ELISA Kit for Antithrombin (AT)

AT3; ATIII; SERPINC1; Anti-Thrombin Antibodies; Serpin Peptidase Inhibitor Clade C Member 1; Coding Sequence Signal Peptide Antithrombin Part 1

Specificity

This assay has high sensitivity and excellent specificity for detection of Antithrombin (AT).
No significant cross-reactivity or interference between Antithrombin (AT) and analogues was observed.

Recovery

Matrices listed below were spiked with certain level of recombinant Antithrombin (AT) and the recovery rates were calculated by comparing the measured value to the expected amount of Antithrombin (AT) in samples.

Matrix Recovery range (%) Average(%)
sodium citrate plasma(n=5) 82-92 87

Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Antithrombin (AT) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Antithrombin (AT) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Antithrombin (AT) and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Sample 1:2 1:4 1:8 1:16
sodium citrate plasma(n=5) 79-90% 88-101% 94-101% 80-93%

Stability

The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

Reagents and materials provided

Reagents Quantity Reagents Quantity
Pre-coated, ready to use 96-well strip plate 1 Plate sealer for 96 wells 4
Standard 2 Standard Diluent 1×20mL
Detection Reagent A 1×120µL Assay Diluent A 1×12mL
Detection Reagent B 1×120µL Assay Diluent B 1×12mL
TMB Substrate 1×9mL Stop Solution 1×6mL
Wash Buffer (30 × concentrate) 1×20mL Instruction manual 1

Assay procedure summary

1. Prepare all reagents, samples and standards;
2. Add 100µL standard or sample to each well. Incubate 1 hours at 37°C;
3. Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C;
4. Aspirate and wash 3 times;
5. Add 100µL prepared Detection Reagent B. Incubate 30 minutes at 37°C;
6. Aspirate and wash 5 times;
7. Add 90µL Substrate Solution. Incubate 10-20 minutes at 37°C;
8. Add 50µL Stop Solution. Read at 450nm immediately.

GIVEAWAYS

INCREMENT SERVICES

Magazine Citations
Thromb Res Thalidomide and multiple myeloma serum synergistically induce a hemostatic imbalance in endothelial cells in vitro PubMed: 25840743
Neurotherapeutics iTRAQ-Based Quantitative Proteomics Reveals the New Evidence Base for Traumatic Brain Injury Treated with Targeted Temperature Management pubmed:29247448
Journal of Clinical and Aesthetic Dermatology Detection of Angiotensin II and AT1 Receptor Concentrations in Keloid and Hypertrophic Scar.
博士论文 An Agile Approach to Novel Recombinant Antithrombin Production for use in Biologics Development
Catalog No. Related products for research use of Homo sapiens (Human) Organism species Applications (RESEARCH USE ONLY!)
NPC313Hu01 Native Antithrombin (AT) Positive Control; Immunogen; SDS-PAGE; WB.
APC313Hu01 Active Antithrombin (AT) Cell culture; Activity Assays.
RPC313Hu03 Recombinant Antithrombin (AT) Positive Control; Immunogen; SDS-PAGE; WB.
EPC313Hu61 Eukaryotic Antithrombin (AT) Positive Control; Immunogen; SDS-PAGE; WB.
RPC313Hu01 Recombinant Antithrombin (AT) Positive Control; Immunogen; SDS-PAGE; WB.
RPC313Hu02 Recombinant Antithrombin (AT) Positive Control; Immunogen; SDS-PAGE; WB.
APC313Hu61 Active Antithrombin (AT) Cell culture; Activity Assays.
PAC313Hu01 Polyclonal Antibody to Antithrombin (AT) WB; IHC; ICC; IP.
PAC313Hu02 Polyclonal Antibody to Antithrombin (AT) WB; IHC; ICC; IP.
PAC313Hu03 Polyclonal Antibody to Antithrombin (AT) WB; IHC; ICC; IP.
LAC313Hu81 FITC-Linked Polyclonal Antibody to Antithrombin (AT) WB; IHC; ICC; IF.
LAC313Hu71 Biotin-Linked Polyclonal Antibody to Antithrombin (AT) WB; IHC; ICC.
MAC313Hu21 Monoclonal Antibody to Antithrombin (AT) WB; IHC; ICC; IP.
MAC313Hu27 Monoclonal Antibody to Antithrombin (AT) WB,IHC
MAC313Hu29 Monoclonal Antibody to Antithrombin (AT) WB; IHC; ICC; IP.
MAC313Hu23 Monoclonal Antibody to Antithrombin (AT) WB; IHC; ICC; IP.
MAC313Hu24 Monoclonal Antibody to Antithrombin (AT) WB; IHC; ICC; IP.
MAC313Hu26 Monoclonal Antibody to Antithrombin (AT) WB; IHC; ICC; IP.
MAC313Hu22 Monoclonal Antibody to Antithrombin (AT) WB,IHC
MAC313Hu28 Monoclonal Antibody to Antithrombin (AT) WB; IHC; ICC; IP.
MAC313Hu25 Monoclonal Antibody to Antithrombin (AT) WB; IHC; ICC; IP.
FAC313Hu02 Anti-Antithrombin (AT) Monoclonal Antibody FCM
FAC313Hu04 Anti-Antithrombin (AT) Monoclonal Antibody FCM
FAC313Hu08 Anti-Antithrombin (AT) Monoclonal Antibody FCM
LAC313Hu72 Biotin-Linked Monoclonal Antibody to Antithrombin (AT) WB; IHC; ICC.
LAC313Hu82 FITC-Linked Monoclonal Antibody to Antithrombin (AT) WB; IHC; ICC; IF.
SEC313Hu ELISA Kit for Antithrombin (AT) Enzyme-linked immunosorbent assay for Antigen Detection.
SCC313Hu CLIA Kit for Antithrombin (AT) Chemiluminescent immunoassay for Antigen Detection.
LMC313Hu Multiplex Assay Kit for Antithrombin (AT) ,etc. by FLIA (Flow Luminescence Immunoassay) FLIA Kit for Antigen Detection.
KSC313Hu01 ELISA Kit DIY Materials for Antithrombin (AT) Main materials for "Do It (ELISA Kit) Yourself".