Multiplex Assay Kit for Spondin 2 (SPON2) ,etc. by CBA (Cytometric Bead Array)

DIL1; Mindin; M-spondin; Extracellular Matrix Protein; Differentially expressed in cancerous and non-cancerous lung cells 1

Specificity

This assay has high sensitivity and excellent specificity for detection of Spondin 2 (SPON2) ,etc. by CBA (Cytometric Bead Array).
No significant cross-reactivity or interference between Spondin 2 (SPON2) ,etc. by CBA (Cytometric Bead Array) and analogues was observed.

Recovery

Matrices listed below were spiked with certain level of recombinant Spondin 2 (SPON2) ,etc. by CBA (Cytometric Bead Array) and the recovery rates were calculated by comparing the measured value to the expected amount of Spondin 2 (SPON2) ,etc. by CBA (Cytometric Bead Array) in samples.

Matrix Recovery range (%) Average(%)
serum(n=5) 92-101 97
EDTA plasma(n=5) 93-104 101
heparin plasma(n=5) 92-101 95

Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Spondin 2 (SPON2) ,etc. by CBA (Cytometric Bead Array) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Spondin 2 (SPON2) ,etc. by CBA (Cytometric Bead Array) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Spondin 2 (SPON2) ,etc. by CBA (Cytometric Bead Array) and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Sample 1:2 1:4 1:8 1:16
serum(n=5) 88-99% 81-95% 95-105% 91-99%
EDTA plasma(n=5) 91-98% 78-99% 78-93% 95-103%
heparin plasma(n=5) 89-98% 94-105% 93-101% 91-98%

Stability

The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

Reagents and materials provided

Reagents Quantity Reagents Quantity

Assay procedure summary

1. Add 200μL analysis buffer solution to each well of the plate for pre-wetting;
2. Preparation of standards, reagents and samples before the experiment;
3. Add 100μL standard or sample to each well, add 10μL magnetic beads, and incubate 120 minutes protect from light at 25°C on shaker;
4. Remove liquid on magnetic frame, add 100μL prepared Detection Reagent A. Incubate 60 minutes protect from light at 25°C on shaker;
5. Remove liquid on magnetic frame, add 100μL prepared Detection Reagent B, and incubate 30 minutes protect from light at 25°C on shaker;
6. Wash plate once on magnetic frame;
7. Add 100μL sheath fluid, swirl for 10 minutes, read on the machine.

GIVEAWAYS

INCREMENT SERVICES

Magazine Citations
The Journal of Urology Spondin-2, a secreted extracellular matrix protein, is a novel diagnostic biomarker for prostate cancer Pubmed: 23665271
Oxford Journals SP136EVALUATION OF SERUM SPONDIN 2 LEVELS IN THE DIFFERENT ETIOLOGIES OF GLOMERULAR DISEASES Content: 30
Nephrology (Carlton) Evaluation of Serum Spondin 2 Levels in the Different Stages of Type 2 Diabetic Nephropathy PubMed: 25973958
Oncogene The extracellular matrix protein mindin attenuates colon cancer progression by blocking angiogenesis via Egr-1-mediated regulation pubmed:28991232
Endocrinology Study of human T21 placenta suggests a potential role of mesenchymal spondin-2 in placental vascular development Pubmed: 30715257
Journal of Cellular and Molecular Medicine Mindin serves as a tumour suppressor gene during colon cancer progression through MAPK/ERK signalling pathway in mice Pubmed: 32614521
Catalog No. Related products for research use of Rattus norvegicus (Rat) Organism species Applications (RESEARCH USE ONLY!)
RPF396Ra01 Recombinant Spondin 2 (SPON2) Positive Control; Immunogen; SDS-PAGE; WB.
PAF396Ra01 Polyclonal Antibody to Spondin 2 (SPON2) IHC
MAF396Ra21 Monoclonal Antibody to Spondin 2 (SPON2) WB; IHC; ICC; IP.
SEF396Ra ELISA Kit for Spondin 2 (SPON2) Enzyme-linked immunosorbent assay for Antigen Detection.
LMF396Ra Multiplex Assay Kit for Spondin 2 (SPON2) ,etc. by FLIA (Flow Luminescence Immunoassay) FLIA Kit for Antigen Detection.
CMF396Ra Multiplex Assay Kit for Spondin 2 (SPON2) ,etc. by CBA (Cytometric Bead Array) FLIA Kit for Antigen Detection.