ELISA Kit for Cystatin C (Cys-C)

CST3; Cystatin 3; Gamma Trace; Post-Gamma-Globulin; Neuroendocrine Basic Polypeptide; Amyloid Angiopathy And Cerebral Hemorrhage

Specificity

This assay has high sensitivity and excellent specificity for detection of Cystatin C (Cys-C).
No significant cross-reactivity or interference between Cystatin C (Cys-C) and analogues was observed.

Recovery

Matrices listed below were spiked with certain level of recombinant Cystatin C (Cys-C) and the recovery rates were calculated by comparing the measured value to the expected amount of Cystatin C (Cys-C) in samples.

Matrix Recovery range (%) Average(%)
serum(n=5) 80-93 90
EDTA plasma(n=5) 85-103 98
heparin plasma(n=5) 79-91 87

Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Cystatin C (Cys-C) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Cystatin C (Cys-C) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Cystatin C (Cys-C) and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Sample 1:2 1:4 1:8 1:16
serum(n=5) 96-103% 80-103% 82-93% 88-95%
EDTA plasma(n=5) 96-103% 93-105% 81-104% 82-97%
heparin plasma(n=5) 86-101% 83-97% 92-103% 90-99%

Stability

The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

Reagents and materials provided

Reagents Quantity Reagents Quantity
Pre-coated, ready to use 96-well strip plate 1 Plate sealer for 96 wells 4
Standard 2 Standard Diluent 1×20mL
Detection Reagent A 1×120µL Assay Diluent A 1×12mL
Detection Reagent B 1×120µL Assay Diluent B 1×12mL
TMB Substrate 1×9mL Stop Solution 1×6mL
Wash Buffer (30 × concentrate) 1×20mL Instruction manual 1

Assay procedure summary

1. Prepare all reagents, samples and standards;
2. Add 100µL standard or sample to each well. Incubate 1 hours at 37°C;
3. Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C;
4. Aspirate and wash 3 times;
5. Add 100µL prepared Detection Reagent B. Incubate 30 minutes at 37°C;
6. Aspirate and wash 5 times;
7. Add 90µL Substrate Solution. Incubate 10-20 minutes at 37°C;
8. Add 50µL Stop Solution. Read at 450nm immediately.

GIVEAWAYS

INCREMENT SERVICES

Magazine Citations
BioMed Research International The Renal Effects of Vanadate Exposure: Potential Biomarkers and Oxidative Stress as a Mechanism of Functional Renal Disorders—Preliminary Studies Hindawi: 740105
Critical Care Endothelial dysfunction and renal fibrosis in endotoxemia-induced oliguric kidney injury: possible role of LPS-binding protein Pubmed:25261195
РОЛЬ АПОПТОЗ ИНДУЦИРУЮЩЕГО ФАКТОРА И ОКИСЛИТЕЛЬНОГО СТРЕССА ПРИ ХРОНИЧЕСКОЙ СЕРДЕЧНОЙ НЕДОСТАТОЧНОСТИ 616-005.8
Journal of Agricultural and Food Chemistry Protective Effects of DHA-PC against Vancomycin-Induced Nephrotoxicity through the Inhibition of Oxidative Stress and Apoptosis in BALB/c Mice pubmed:29254330
THERAPEUTIC ADVANCES IN RESPIRATORY DISEASE Association of cathepsin B and cystatin C with an age-related pulmonary subclinical state in a healthy Chinese population Pubmed: 32401159
RSC Adv SERS based Y-shaped aptasensor for early diagnosis of acute kidney injury Pubmed:35733690
Food Funct A Compared Study of Gentle Ketogenic Diet Containing Medium-Chain Triglycerides or Long-Chain Triglycerides on Chronic Sleep Deprivation-Induced Cognitive … Pubmed:35141738
Catalog No. Related products for research use of Oryctolagus cuniculus (Rabbit) Organism species Applications (RESEARCH USE ONLY!)
SEA896Rb ELISA Kit for Cystatin C (Cys-C) Enzyme-linked immunosorbent assay for Antigen Detection.
LMA896Rb Multiplex Assay Kit for Cystatin C (Cys-C) ,etc. by FLIA (Flow Luminescence Immunoassay) FLIA Kit for Antigen Detection.