ELISA Kit for Slit Homolog 2 (Slit2)

SLIL3

Specificity

This assay has high sensitivity and excellent specificity for detection of Slit Homolog 2 (Slit2).
No significant cross-reactivity or interference between Slit Homolog 2 (Slit2) and analogues was observed.

Recovery

Matrices listed below were spiked with certain level of recombinant Slit Homolog 2 (Slit2) and the recovery rates were calculated by comparing the measured value to the expected amount of Slit Homolog 2 (Slit2) in samples.

Matrix Recovery range (%) Average(%)
serum(n=5) 79-96 91
EDTA plasma(n=5) 90-98 94
heparin plasma(n=5) 98-105 101

Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Slit Homolog 2 (Slit2) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Slit Homolog 2 (Slit2) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Slit Homolog 2 (Slit2) and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Sample 1:2 1:4 1:8 1:16
serum(n=5) 95-103% 88-102% 79-93% 93-105%
EDTA plasma(n=5) 80-92% 98-105% 89-99% 80-101%
heparin plasma(n=5) 99-105% 90-99% 95-105% 81-103%

Stability

The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

Reagents and materials provided

Reagents Quantity Reagents Quantity
Pre-coated, ready to use 96-well strip plate 1 Plate sealer for 96 wells 4
Standard 2 Standard Diluent 1×20mL
Detection Reagent A 1×120µL Assay Diluent A 1×12mL
Detection Reagent B 1×120µL Assay Diluent B 1×12mL
TMB Substrate 1×9mL Stop Solution 1×6mL
Wash Buffer (30 × concentrate) 1×20mL Instruction manual 1

Assay procedure summary

1. Prepare all reagents, samples and standards;
2. Add 100µL standard or sample to each well. Incubate 1 hours at 37°C;
3. Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C;
4. Aspirate and wash 3 times;
5. Add 100µL prepared Detection Reagent B. Incubate 30 minutes at 37°C;
6. Aspirate and wash 5 times;
7. Add 90µL Substrate Solution. Incubate 10-20 minutes at 37°C;
8. Add 50µL Stop Solution. Read at 450nm immediately.

GIVEAWAYS

INCREMENT SERVICES

Magazine Citations
Cancer Research SLIT2 Attenuation during Lung Cancer Progression Deregulates β-Catenin and E-Cadherin and Associates with Poor Prognosis AACR: 702543
Carcinogenesis Slit-2 facilitates interaction of P-cadherin with Robo-3 and inhibits cell migration in an oral squamous cell carcinoma cell line OxfordJournals: source
PLoS One Analytical Performance of ELISA Assays in Urine: One More Bottleneck towards Biomarker Validation and Clinical Implementation Pubmed:26889680
ESC Heart Fail Crosstalk between the activated Slit2¨CRobo1 pathway and TGF©\¦Â1 signalling promotes cardiac fibrosis 33236535
Catalog No. Related products for research use of Rattus norvegicus (Rat) Organism species Applications (RESEARCH USE ONLY!)
APA672Ra01 Active Slit Homolog 2 (Slit2) Cell culture; Activity Assays.
RPA672Ra01 Recombinant Slit Homolog 2 (Slit2) Positive Control; Immunogen; SDS-PAGE; WB.
PAA672Ra01 Polyclonal Antibody to Slit Homolog 2 (Slit2) WB; IHC; ICC; IP.
LAA672Ra71 Biotin-Linked Polyclonal Antibody to Slit Homolog 2 (Slit2) WB; IHC; ICC.
MAA672Ra21 Monoclonal Antibody to Slit Homolog 2 (Slit2) WB; IHC; ICC; IP.
SEA672Ra ELISA Kit for Slit Homolog 2 (Slit2) Enzyme-linked immunosorbent assay for Antigen Detection.
LMA672Ra Multiplex Assay Kit for Slit Homolog 2 (Slit2) ,etc. by FLIA (Flow Luminescence Immunoassay) FLIA Kit for Antigen Detection.