Multiplex Assay Kit for N-Terminal Pro-Atrial Natriuretic Peptide (NT-ProANP) ,etc. by FLIA (Flow Luminescence Immunoassay)

NT-Pro-ANP; N-ANP

(Note: Up to 8-plex in one testing reaction)

Specificity

This assay has high sensitivity and excellent specificity for detection of N-Terminal Pro-Atrial Natriuretic Peptide (NT-ProANP) ,etc. by FLIA (Flow Luminescence Immunoassay).
No significant cross-reactivity or interference between N-Terminal Pro-Atrial Natriuretic Peptide (NT-ProANP) ,etc. by FLIA (Flow Luminescence Immunoassay) and analogues was observed.

Recovery

Matrices listed below were spiked with certain level of recombinant N-Terminal Pro-Atrial Natriuretic Peptide (NT-ProANP) ,etc. by FLIA (Flow Luminescence Immunoassay) and the recovery rates were calculated by comparing the measured value to the expected amount of N-Terminal Pro-Atrial Natriuretic Peptide (NT-ProANP) ,etc. by FLIA (Flow Luminescence Immunoassay) in samples.

Matrix Recovery range (%) Average(%)
serum(n=5) 90-104 97
EDTA plasma(n=5) 78-102 91
heparin plasma(n=5) 84-92 88

Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level N-Terminal Pro-Atrial Natriuretic Peptide (NT-ProANP) ,etc. by FLIA (Flow Luminescence Immunoassay) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level N-Terminal Pro-Atrial Natriuretic Peptide (NT-ProANP) ,etc. by FLIA (Flow Luminescence Immunoassay) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of N-Terminal Pro-Atrial Natriuretic Peptide (NT-ProANP) ,etc. by FLIA (Flow Luminescence Immunoassay) and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Sample 1:2 1:4 1:8 1:16
serum(n=5) 86-101% 80-91% 90-99% 83-90%
EDTA plasma(n=5) 79-101% 96-103% 81-91% 87-95%
heparin plasma(n=5) 86-96% 86-98% 90-102% 78-101%

Stability

The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

Reagents and materials provided

Reagents Quantity Reagents Quantity
96-well plate 1 Plate sealer for 96 wells 4
Pre-Mixed Standard 2 Standard Diluent 1×20mL
Pre-Mixed Magnetic beads (22#:NT-ProANP) 1 Analysis buffer 1×20mL
Pre-Mixed Detection Reagent A 1×120μL Assay Diluent A 1×12mL
Detection Reagent B (PE-SA) 1×120μL Assay Diluent B 1×12mL
Sheath Fluid 1×10mL Wash Buffer (30 × concentrate) 1×20mL
Instruction manual 1

Assay procedure summary

1. Preparation of standards, reagents and samples before the experiment;
2. Add 100μL standard or sample to each well,
    add 10μL magnetic beads, and incubate 90min at 37°C on shaker;
3. Remove liquid on magnetic frame, add 100μL prepared Detection Reagent A. Incubate 60min at 37°C on shaker;
4. Wash plate on magnetic frame for three times;
5. Add 100μL prepared Detection Reagent B, and incubate 30 min at 37°C on shaker;
6. Wash plate on magnetic frame for three times;
7. Add 100μL sheath solution, swirl for 2 minutes, read on the machine.

GIVEAWAYS

INCREMENT SERVICES

Magazine Citations
European Veterinary Conference Atrial natriuretic peptides detect cardiac dilatation Biblio.ugent: 3227757
Echocardiography Prediction of postoperative atrial fibrillation with left atrial mechanical functions and NT‐pro ANP levels after coronary artery bypass surgery: A three‐dimensional … Pubmed:29457260
Циррозы печени вирусной этиологии у детей
Van Veterinary Journal Van Veterinary Journal
Journal of the American Heart Association Recombinant Soluble Corin Improves Cardiac Function in Mouse Models of Heart Failure 33759549
Catalog No. Related products for research use of Rattus norvegicus (Rat) Organism species Applications (RESEARCH USE ONLY!)
RPA484Ra01 Recombinant N-Terminal Pro-Atrial Natriuretic Peptide (NT-ProANP) Positive Control; Immunogen; SDS-PAGE; WB.
RPA484Ra02 Recombinant N-Terminal Pro-Atrial Natriuretic Peptide (NT-ProANP) Positive Control; Immunogen; SDS-PAGE; WB.
PAA484Ra02 Polyclonal Antibody to N-Terminal Pro-Atrial Natriuretic Peptide (NT-ProANP) WB; IHC; ICC; IP.
PAA484Ra01 Polyclonal Antibody to N-Terminal Pro-Atrial Natriuretic Peptide (NT-ProANP) WB; IHC; ICC; IP.
MAA484Ra21 Monoclonal Antibody to N-Terminal Pro-Atrial Natriuretic Peptide (NT-ProANP) WB; IHC; ICC; IP.
SEA484Ra ELISA Kit for N-Terminal Pro-Atrial Natriuretic Peptide (NT-ProANP) Enzyme-linked immunosorbent assay for Antigen Detection.
LMA484Ra Multiplex Assay Kit for N-Terminal Pro-Atrial Natriuretic Peptide (NT-ProANP) ,etc. by FLIA (Flow Luminescence Immunoassay) FLIA Kit for Antigen Detection.
KSA484Ra01 ELISA Kit DIY Materials for N-Terminal Pro-Atrial Natriuretic Peptide (NT-ProANP) Main materials for "Do It(ELISA Kit) Yourself"