Multiplex Assay Kit for Syndecan 4 (SDC4) ,etc. by CBA (Cytometric Bead Array)

SYND4; Amphiglycan; Ryudocan; Ryudocan core protein

Specificity

This assay has high sensitivity and excellent specificity for detection of Syndecan 4 (SDC4) ,etc. by CBA (Cytometric Bead Array).
No significant cross-reactivity or interference between Syndecan 4 (SDC4) ,etc. by CBA (Cytometric Bead Array) and analogues was observed.

Recovery

Matrices listed below were spiked with certain level of recombinant Syndecan 4 (SDC4) ,etc. by CBA (Cytometric Bead Array) and the recovery rates were calculated by comparing the measured value to the expected amount of Syndecan 4 (SDC4) ,etc. by CBA (Cytometric Bead Array) in samples.

Matrix Recovery range (%) Average(%)
serum(n=5) 92-105 98
EDTA plasma(n=5) 79-104 86
heparin plasma(n=5) 82-101 85

Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Syndecan 4 (SDC4) ,etc. by CBA (Cytometric Bead Array) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Syndecan 4 (SDC4) ,etc. by CBA (Cytometric Bead Array) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Syndecan 4 (SDC4) ,etc. by CBA (Cytometric Bead Array) and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Sample 1:2 1:4 1:8 1:16
serum(n=5) 78-96% 82-104% 90-104% 85-93%
EDTA plasma(n=5) 94-105% 84-91% 83-101% 97-104%
heparin plasma(n=5) 95-105% 89-96% 98-105% 96-105%

Stability

The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

Reagents and materials provided

Reagents Quantity Reagents Quantity

Assay procedure summary

1. Add 200μL analysis buffer solution to each well of the plate for pre-wetting;
2. Preparation of standards, reagents and samples before the experiment;
3. Add 100μL standard or sample to each well, add 10μL magnetic beads, and incubate 120 minutes protect from light at 25°C on shaker;
4. Remove liquid on magnetic frame, add 100μL prepared Detection Reagent A. Incubate 60 minutes protect from light at 25°C on shaker;
5. Remove liquid on magnetic frame, add 100μL prepared Detection Reagent B, and incubate 30 minutes protect from light at 25°C on shaker;
6. Wash plate once on magnetic frame;
7. Add 100μL sheath fluid, swirl for 10 minutes, read on the machine.

GIVEAWAYS

INCREMENT SERVICES

Magazine Citations
Lipids in Health and Disease The influence of atorvastatin on parameters of inflammation left ventricular function, hospitalizations and mortality in patients with dilated cardiomyopathy–5-year follow-up PubMed: PMC3641983
The FASEB Journal Matrix metalloproteinase 9-mediated shedding of syndecan 4 in response to tumor necrosis factor α: a contributor to endothelial cell glycocalyx dysfunction Pubmed:25122554
Scientific Reports Biomarkers and Echocardiographic Predictors of Myocardial Dysfunction in Patients with Hypertension PubMed: 25747153
Int J Mol Sci The Multi-Biomarker Approach for Heart Failure in Patients with Hypertension PubMed: 25984599
Biochemical and Biophysical Research Communications Syndecan-4 shedding impairs macrovascular angiogenesis in diabetes mellitus Pubmed:27018253
Kidney International Aldosterone induces albuminuria via matrix metalloproteinase–dependent damage of the endothelial glycocalyx Pubmed: 30389198
Catalog No. Related products for research use of Homo sapiens (Human) Organism species Applications (RESEARCH USE ONLY!)
APB939Hu01 Active Syndecan 4 (SDC4) Cell culture; Activity Assays.
EPB939Hu61 Eukaryotic Syndecan 4 (SDC4) Positive Control; Immunogen; SDS-PAGE; WB.
RPB939Hu01 Recombinant Syndecan 4 (SDC4) Positive Control; Immunogen; SDS-PAGE; WB.
PAB939Hu01 Polyclonal Antibody to Syndecan 4 (SDC4) WB
MAB939Hu22 Monoclonal Antibody to Syndecan 4 (SDC4) Testing in progress
SEB939Hu ELISA Kit for Syndecan 4 (SDC4) Enzyme-linked immunosorbent assay for Antigen Detection.
CMB939Hu Multiplex Assay Kit for Syndecan 4 (SDC4) ,etc. by CBA (Cytometric Bead Array) FLIA Kit for Antigen Detection.
LMB939Hu Multiplex Assay Kit for Syndecan 4 (SDC4) ,etc. by FLIA (Flow Luminescence Immunoassay) FLIA Kit for Antigen Detection.