Multiplex Assay Kit for Low Density Lipoprotein Receptor (LDLR) ,etc. by CBA (Cytometric Bead Array)

FH; FHC; Familial Hypercholesterolemia

Specificity

This assay has high sensitivity and excellent specificity for detection of Low Density Lipoprotein Receptor (LDLR) ,etc. by CBA (Cytometric Bead Array).
No significant cross-reactivity or interference between Low Density Lipoprotein Receptor (LDLR) ,etc. by CBA (Cytometric Bead Array) and analogues was observed.

Recovery

Matrices listed below were spiked with certain level of recombinant Low Density Lipoprotein Receptor (LDLR) ,etc. by CBA (Cytometric Bead Array) and the recovery rates were calculated by comparing the measured value to the expected amount of Low Density Lipoprotein Receptor (LDLR) ,etc. by CBA (Cytometric Bead Array) in samples.

Matrix Recovery range (%) Average(%)
serum(n=5) 78-101 92
EDTA plasma(n=5) 94-102 98
heparin plasma(n=5) 87-95 90

Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Low Density Lipoprotein Receptor (LDLR) ,etc. by CBA (Cytometric Bead Array) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Low Density Lipoprotein Receptor (LDLR) ,etc. by CBA (Cytometric Bead Array) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Low Density Lipoprotein Receptor (LDLR) ,etc. by CBA (Cytometric Bead Array) and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Sample 1:2 1:4 1:8 1:16
serum(n=5) 79-97% 83-97% 95-104% 97-104%
EDTA plasma(n=5) 94-102% 91-101% 82-90% 89-102%
heparin plasma(n=5) 97-105% 80-99% 94-103% 88-95%

Stability

The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

Reagents and materials provided

Reagents Quantity Reagents Quantity

Assay procedure summary

1. Add 200μL analysis buffer solution to each well of the plate for pre-wetting;
2. Preparation of standards, reagents and samples before the experiment;
3. Add 100μL standard or sample to each well, add 10μL magnetic beads, and incubate 120 minutes protect from light at 25°C on shaker;
4. Remove liquid on magnetic frame, add 100μL prepared Detection Reagent A. Incubate 60 minutes protect from light at 25°C on shaker;
5. Remove liquid on magnetic frame, add 100μL prepared Detection Reagent B, and incubate 30 minutes protect from light at 25°C on shaker;
6. Wash plate once on magnetic frame;
7. Add 100μL sheath fluid, swirl for 10 minutes, read on the machine.

GIVEAWAYS

INCREMENT SERVICES

Magazine Citations
Neurotoxicology. 27-Hydroxycholesterol regulates cholesterol synthesis and transport in C6 glioma cells. pubmed:28167099
Oncotarget Synthetic low-density lipoprotein (sLDL) selectively delivers paclitaxel to tumor with lowsystemic toxicity. pubmed:27409176
Catalog No. Related products for research use of Homo sapiens (Human) Organism species Applications (RESEARCH USE ONLY!)
RPB008Hu01 Recombinant Low Density Lipoprotein Receptor (LDLR) Positive Control; Immunogen; SDS-PAGE; WB.
PAB008Hu01 Polyclonal Antibody to Low Density Lipoprotein Receptor (LDLR) IHC; FCM
FAB008Hu01 Anti-Low Density Lipoprotein Receptor (LDLR) Polyclonal Antibody FCM
MAB008Hu22 Monoclonal Antibody to Low Density Lipoprotein Receptor (LDLR) WB; IHC; ICC; IP.
SEB008Hu ELISA Kit for Low Density Lipoprotein Receptor (LDLR) Enzyme-linked immunosorbent assay for Antigen Detection.
LMB008Hu Multiplex Assay Kit for Low Density Lipoprotein Receptor (LDLR) ,etc. by FLIA (Flow Luminescence Immunoassay) FLIA Kit for Antigen Detection.
CMB008Hu Multiplex Assay Kit for Low Density Lipoprotein Receptor (LDLR) ,etc. by CBA (Cytometric Bead Array) FLIA Kit for Antigen Detection.