Multiplex Assay Kit for Coagulation Factor IX (F9) ,etc. by CBA (Cytometric Bead Array)

HEMB; FIX; GLA Domain; PTC; Anti Hemophilic Factor B; Christmas Factor; Plasma Thromboplastic Component; Christmas Disease; Hemophilia B

Specificity

This assay has high sensitivity and excellent specificity for detection of Coagulation Factor IX (F9) ,etc. by CBA (Cytometric Bead Array).
No significant cross-reactivity or interference between Coagulation Factor IX (F9) ,etc. by CBA (Cytometric Bead Array) and analogues was observed.

Recovery

Matrices listed below were spiked with certain level of recombinant Coagulation Factor IX (F9) ,etc. by CBA (Cytometric Bead Array) and the recovery rates were calculated by comparing the measured value to the expected amount of Coagulation Factor IX (F9) ,etc. by CBA (Cytometric Bead Array) in samples.

Matrix Recovery range (%) Average(%)
serum(n=5) 78-101 84
EDTA plasma(n=5) 80-99 93
heparin plasma(n=5) 78-99 90

Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Coagulation Factor IX (F9) ,etc. by CBA (Cytometric Bead Array) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Coagulation Factor IX (F9) ,etc. by CBA (Cytometric Bead Array) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Coagulation Factor IX (F9) ,etc. by CBA (Cytometric Bead Array) and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Sample 1:2 1:4 1:8 1:16
serum(n=5) 95-102% 86-101% 80-90% 95-103%
EDTA plasma(n=5) 82-104% 85-96% 93-103% 96-105%
heparin plasma(n=5) 80-99% 86-94% 85-99% 82-99%

Stability

The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

Reagents and materials provided

Reagents Quantity Reagents Quantity

Assay procedure summary

1. Add 200μL analysis buffer solution to each well of the plate for pre-wetting;
2. Preparation of standards, reagents and samples before the experiment;
3. Add 100μL standard or sample to each well, add 10μL magnetic beads, and incubate 120 minutes protect from light at 25°C on shaker;
4. Remove liquid on magnetic frame, add 100μL prepared Detection Reagent A. Incubate 60 minutes protect from light at 25°C on shaker;
5. Remove liquid on magnetic frame, add 100μL prepared Detection Reagent B, and incubate 30 minutes protect from light at 25°C on shaker;
6. Wash plate once on magnetic frame;
7. Add 100μL sheath fluid, swirl for 10 minutes, read on the machine.

GIVEAWAYS

INCREMENT SERVICES

Magazine Citations
Cardiovascular Research Loss of accelerates arterial thrombosis by increasing formation of neutrophil extracellular traps and plasma tissue factor activity Pubmed:29444200
Catalog No. Related products for research use of Homo sapiens (Human) Organism species Applications (RESEARCH USE ONLY!)
RPA840Hu01 Recombinant Coagulation Factor IX (F9) Positive Control; Immunogen; SDS-PAGE; WB.
PAA840Hu01 Polyclonal Antibody to Coagulation Factor IX (F9) WB
LAA840Hu91 HRP-Linked Polyclonal Antibody to Coagulation Factor IX (F9) WB
MAA840Hu22 Monoclonal Antibody to Coagulation Factor IX (F9) WB
MAA840Hu23 Monoclonal Antibody to Coagulation Factor IX (F9) ICC/IF
MAA840Hu21 Monoclonal Antibody to Coagulation Factor IX (F9) WB
FAA840Hu02 Anti-Coagulation Factor IX (F9) Monoclonal Antibody FCM
FAA840Hu04 Anti-Coagulation Factor IX (F9) Monoclonal Antibody FCM
SEA840Hu ELISA Kit for Coagulation Factor IX (F9) Enzyme-linked immunosorbent assay for Antigen Detection.
LMA840Hu Multiplex Assay Kit for Coagulation Factor IX (F9) ,etc. by FLIA (Flow Luminescence Immunoassay) FLIA Kit for Antigen Detection.
CMA840Hu Multiplex Assay Kit for Coagulation Factor IX (F9) ,etc. by CBA (Cytometric Bead Array) FLIA Kit for Antigen Detection.