ELISA Kit for A Disintegrin And Metalloproteinase With Thrombospondin 5 (ADAMTS5)

ADMP-2; Aggrecanase-2

Specificity

This assay has high sensitivity and excellent specificity for detection of A Disintegrin And Metalloproteinase With Thrombospondin 5 (ADAMTS5).
No significant cross-reactivity or interference between A Disintegrin And Metalloproteinase With Thrombospondin 5 (ADAMTS5) and analogues was observed.

Recovery

Matrices listed below were spiked with certain level of recombinant A Disintegrin And Metalloproteinase With Thrombospondin 5 (ADAMTS5) and the recovery rates were calculated by comparing the measured value to the expected amount of A Disintegrin And Metalloproteinase With Thrombospondin 5 (ADAMTS5) in samples.

Matrix Recovery range (%) Average(%)
serum(n=5) 89-97 92
EDTA plasma(n=5) 93-101 96
heparin plasma(n=5) 86-96 92

Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level A Disintegrin And Metalloproteinase With Thrombospondin 5 (ADAMTS5) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level A Disintegrin And Metalloproteinase With Thrombospondin 5 (ADAMTS5) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of A Disintegrin And Metalloproteinase With Thrombospondin 5 (ADAMTS5) and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Sample 1:2 1:4 1:8 1:16
serum(n=5) 84-91% 90-97% 92-105% 95-102%
EDTA plasma(n=5) 79-96% 93-104% 86-104% 94-101%
heparin plasma(n=5) 83-99% 94-101% 96-103% 87-94%

Stability

The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

Reagents and materials provided

Reagents Quantity Reagents Quantity
Pre-coated, ready to use 96-well strip plate 1 Plate sealer for 96 wells 4
Standard 2 Standard Diluent 1×20mL
Detection Reagent A 1×120µL Assay Diluent A 1×12mL
Detection Reagent B 1×120µL Assay Diluent B 1×12mL
TMB Substrate 1×9mL Stop Solution 1×6mL
Wash Buffer (30 × concentrate) 1×20mL Instruction manual 1

Assay procedure summary

1. Prepare all reagents, samples and standards;
2. Add 100µL standard or sample to each well. Incubate 1 hours at 37°C;
3. Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C;
4. Aspirate and wash 3 times;
5. Add 100µL prepared Detection Reagent B. Incubate 30 minutes at 37°C;
6. Aspirate and wash 5 times;
7. Add 90µL Substrate Solution. Incubate 10-20 minutes at 37°C;
8. Add 50µL Stop Solution. Read at 450nm immediately.

GIVEAWAYS

INCREMENT SERVICES

Magazine Citations
PLoS ONE A Combinatorial Relative Mass Value Evaluation of Endogenous Bioactive Proteins in Three-Dimensional Cultured Nucleus Pulposus Cells of Herniated Intervertebral Discs: Identification of Potential Target Proteins for Gene Therapeutic Approaches Plosone: Source
Osteoarthritis and Cartilage Lack of anti-inflammatory and anti-catabolic effects on basal inflamed osteoarthritic chondrocytes or synoviocytes by adipose stem cell-conditioned medium PubMed: 26521751
Journal of Cellular and Molecular Medicine VIP and CRF reduce ADAMTS expression and function in osteoarthritis synovial fibroblasts Pubmed:26818776
Int Wound J. From varices to venous ulceration: the story of chronic venous disease described by metalloproteinases. Pubmed:26991748
The American journal of pathology Healthy and Osteoarthritic Synovial Fibroblasts Produce a Disintegrin and Metalloproteinase with Thrombospondin Motifs 4, 5, 7, and 12: Induction by IL-1β and Fibronectin and Contribution to Cartilage Damage. pubmed:27449198
PLoS One. Self-complementary adeno-associated virus serotype 6 mediated knockdown of ADAMTS4 induces long-term and effective enhancement of aggrecan in degenerative human nucleus pulposus cells: A new therapeutic approach for intervertebral disc disorders. pubmed:28207788
PLoS One Self-complementary adeno-associated virus serotype 6 mediated knockdown of ADAMTS4 induces long-term and effective enhancement of aggrecan in degenerative human nucleus pulposus cells: A new therapeutic approach for intervertebral disc disorders 10.1371:journal.pone.0172181
International Wound Journal From varices to venous ulceration: the story of chronic venous disease described by metalloproteinases pubmed:26991748
BMC musculoskeletal disorders Hemoglobin stimulates the expression of ADAMTS-5 and ADAMTS-9 by synovial cells: a possible cause of articular cartilage damage after intra-articular hemorrhage. pubmed:29137610
Journal of Clinical Medicine The Effects of TNF-α Inhibition on the Metabolism of Cartilage: Relationship between KS, HA, HAPLN1 and ADAMTS4, ADAMTS5, TOS and TGF-β1 Plasma … Pubmed:35407621
Catalog No. Related products for research use of Rattus norvegicus (Rat) Organism species Applications (RESEARCH USE ONLY!)
RPK205Ra01 Recombinant A Disintegrin And Metalloproteinase With Thrombospondin 5 (ADAMTS5) Positive Control; Immunogen; SDS-PAGE; WB.
PAK205Ra01 Polyclonal Antibody to A Disintegrin And Metalloproteinase With Thrombospondin 5 (ADAMTS5) WB
MAK205Ra21 Monoclonal Antibody to A Disintegrin And Metalloproteinase With Thrombospondin 5 (ADAMTS5) WB; IHC; ICC; IP.
SEK205Ra ELISA Kit for A Disintegrin And Metalloproteinase With Thrombospondin 5 (ADAMTS5) Enzyme-linked immunosorbent assay for Antigen Detection.
LMK205Ra Multiplex Assay Kit for A Disintegrin And Metalloproteinase With Thrombospondin 5 (ADAMTS5) ,etc. by FLIA (Flow Luminescence Immunoassay) FLIA Kit for Antigen Detection.