ELISA Kit for Uromodulin (UMOD)

THP; Uromucoid,Tamm-Horsfall Glycoprotein; Tamm-Horsfall urinary glycoprotein

Specificity

This assay has high sensitivity and excellent specificity for detection of Uromodulin (UMOD).
No significant cross-reactivity or interference between Uromodulin (UMOD) and analogues was observed.

Recovery

Matrices listed below were spiked with certain level of recombinant Uromodulin (UMOD) and the recovery rates were calculated by comparing the measured value to the expected amount of Uromodulin (UMOD) in samples.

Matrix Recovery range (%) Average(%)
serum(n=5) 88-104 97
EDTA plasma(n=5) 79-104 85
heparin plasma(n=5) 88-95 92

Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Uromodulin (UMOD) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Uromodulin (UMOD) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Uromodulin (UMOD) and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Sample 1:2 1:4 1:8 1:16
serum(n=5) 94-101% 79-98% 80-101% 84-92%
EDTA plasma(n=5) 78-90% 79-89% 78-90% 87-95%
heparin plasma(n=5) 78-90% 97-105% 83-91% 96-104%

Stability

The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

Reagents and materials provided

Reagents Quantity Reagents Quantity
Pre-coated, ready to use 96-well strip plate 1 Plate sealer for 96 wells 4
Standard 2 Standard Diluent 1×20mL
Detection Reagent A 1×120µL Assay Diluent A 1×12mL
Detection Reagent B 1×120µL Assay Diluent B 1×12mL
TMB Substrate 1×9mL Stop Solution 1×6mL
Wash Buffer (30 × concentrate) 1×20mL Instruction manual 1

Assay procedure summary

1. Prepare all reagents, samples and standards;
2. Add 100µL standard or sample to each well. Incubate 1 hours at 37°C;
3. Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C;
4. Aspirate and wash 3 times;
5. Add 100µL prepared Detection Reagent B. Incubate 30 minutes at 37°C;
6. Aspirate and wash 5 times;
7. Add 90µL Substrate Solution. Incubate 10-20 minutes at 37°C;
8. Add 50µL Stop Solution. Read at 450nm immediately.

GIVEAWAYS

INCREMENT SERVICES

Magazine Citations
Evid Based Complement Alternat Med Quercetin Protects against Cadmium-Induced Renal Uric Acid Transport System Alteration and Lipid Metabolism Disorder in Rats PubMed: PMC3368504
Journal of Korean Medical Science Fate of Neutrophils during the Recovery Phase of Ischemia/Reperfusion Induced Acute Kidney Injury pubmed:28875605
Eberhard Karls University Experimental nephrotic syndrome in mice with an inducible deletion of podocin and the role of plasminogen in sodium retention
Catalog No. Related products for research use of Mus musculus (Mouse) Organism species Applications (RESEARCH USE ONLY!)
RPG918Mu02 Recombinant Uromodulin (UMOD) Positive Control; Immunogen; SDS-PAGE; WB.
RPG918Mu01 Recombinant Uromodulin (UMOD) Positive Control; Immunogen; SDS-PAGE; WB.
PAG918Mu01 Polyclonal Antibody to Uromodulin (UMOD) WB; IHC; ICC; IP.
PAG918Mu02 Polyclonal Antibody to Uromodulin (UMOD) WB; IHC; ICC; IP.
SEG918Mu ELISA Kit for Uromodulin (UMOD) Enzyme-linked immunosorbent assay for Antigen Detection.
LMG918Mu Multiplex Assay Kit for Uromodulin (UMOD) ,etc. by FLIA (Flow Luminescence Immunoassay) FLIA Kit for Antigen Detection.