ELISA Kit for Fibrinogen Gamma (FGg)

FG-G

Specificity

This assay has high sensitivity and excellent specificity for detection of Fibrinogen Gamma (FGg).
No significant cross-reactivity or interference between Fibrinogen Gamma (FGg) and analogues was observed.

Recovery

Matrices listed below were spiked with certain level of recombinant Fibrinogen Gamma (FGg) and the recovery rates were calculated by comparing the measured value to the expected amount of Fibrinogen Gamma (FGg) in samples.

Matrix Recovery range (%) Average(%)
serum(n=5) 85-99 95
EDTA plasma(n=5) 78-101 87
heparin plasma(n=5) 89-97 94

Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Fibrinogen Gamma (FGg) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Fibrinogen Gamma (FGg) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Fibrinogen Gamma (FGg) and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Sample 1:2 1:4 1:8 1:16
serum(n=5) 97-105% 92-105% 91-105% 91-99%
EDTA plasma(n=5) 78-91% 97-105% 94-105% 86-93%
heparin plasma(n=5) 88-104% 93-101% 82-98% 79-93%

Stability

The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

Reagents and materials provided

Reagents Quantity Reagents Quantity
Pre-coated, ready to use 96-well strip plate 1 Plate sealer for 96 wells 4
Standard 2 Standard Diluent 1×20mL
Detection Reagent A 1×120µL Assay Diluent A 1×12mL
Detection Reagent B 1×120µL Assay Diluent B 1×12mL
TMB Substrate 1×9mL Stop Solution 1×6mL
Wash Buffer (30 × concentrate) 1×20mL Instruction manual 1

Assay procedure summary

1. Prepare all reagents, samples and standards;
2. Add 100µL standard or sample to each well. Incubate 1 hours at 37°C;
3. Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C;
4. Aspirate and wash 3 times;
5. Add 100µL prepared Detection Reagent B. Incubate 30 minutes at 37°C;
6. Aspirate and wash 5 times;
7. Add 90µL Substrate Solution. Incubate 10-20 minutes at 37°C;
8. Add 50µL Stop Solution. Read at 450nm immediately.

GIVEAWAYS

INCREMENT SERVICES

Magazine Citations
Neurotherapeutics iTRAQ-Based Quantitative Proteomics Reveals the New Evidence Base for Traumatic Brain Injury Treated with Targeted Temperature Management pubmed:29247448
Alzheimer's Research & Therapy The influence of insulin resistance on cerebrospinal fluid and plasma biomarkers of Alzheimer’s pathology pubmed:28441961
GLYCOCONJUGATE JOURNAL Nattokinase-heparin exhibits beneficial efficacy and safety—an optimal strategy for CKD patients on hemodialysis Pubmed: 30788657
JOURNAL OF CONTROLLED RELEASE Polycaprolactone vascular graft with epigallocatechin gallate embedded sandwiched layer-by-layer functionalization for enhanced antithrombogenicity and anti … Pubmed: 31982435
Catalog No. Related products for research use of Rattus norvegicus (Rat) Organism species Applications (RESEARCH USE ONLY!)
RPC477Ra01 Recombinant Fibrinogen Gamma (FGg) Positive Control; Immunogen; SDS-PAGE; WB.
PAC477Ra01 Polyclonal Antibody to Fibrinogen Gamma (FGg) WB; IHC; ICC; IP.
LAC477Ra71 Biotin-Linked Polyclonal Antibody to Fibrinogen Gamma (FGg) WB; IHC; ICC.
LAC477Ra81 FITC-Linked Polyclonal Antibody to Fibrinogen Gamma (FGg) WB; IHC; ICC; IF.
MAC477Ra21 Monoclonal Antibody to Fibrinogen Gamma (FGg) WB; IHC; ICC; IP.
SEC477Ra ELISA Kit for Fibrinogen Gamma (FGg) Enzyme-linked immunosorbent assay for Antigen Detection.
LMC477Ra Multiplex Assay Kit for Fibrinogen Gamma (FGg) ,etc. by FLIA (Flow Luminescence Immunoassay) FLIA Kit for Antigen Detection.
KSC477Ra01 ELISA Kit DIY Materials for Fibrinogen Gamma (FGg) Main materials for "Do It (ELISA Kit) Yourself".