ELISA Kit for Complement Component 9 (C9)

Specificity

This assay has high sensitivity and excellent specificity for detection of Complement Component 9 (C9).
No significant cross-reactivity or interference between Complement Component 9 (C9) and analogues was observed.

Recovery

Matrices listed below were spiked with certain level of recombinant Complement Component 9 (C9) and the recovery rates were calculated by comparing the measured value to the expected amount of Complement Component 9 (C9) in samples.

Matrix Recovery range (%) Average(%)
serum(n=5) 87-101 92
EDTA plasma(n=5) 84-101 95
heparin plasma(n=5) 85-93 90

Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Complement Component 9 (C9) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Complement Component 9 (C9) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Complement Component 9 (C9) and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Sample 1:2 1:4 1:8 1:16
serum(n=5) 78-91% 91-99% 80-101% 98-105%
EDTA plasma(n=5) 87-98% 78-95% 79-89% 96-103%
heparin plasma(n=5) 80-102% 94-101% 86-103% 87-103%

Stability

The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

Reagents and materials provided

Reagents Quantity Reagents Quantity
Pre-coated, ready to use 96-well strip plate 1 Plate sealer for 96 wells 4
Standard 2 Standard Diluent 1×20mL
Detection Reagent A 1×120µL Assay Diluent A 1×12mL
Detection Reagent B 1×120µL Assay Diluent B 1×12mL
TMB Substrate 1×9mL Stop Solution 1×6mL
Wash Buffer (30 × concentrate) 1×20mL Instruction manual 1

Assay procedure summary

1. Prepare all reagents, samples and standards;
2. Add 100µL standard or sample to each well. Incubate 1 hours at 37°C;
3. Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C;
4. Aspirate and wash 3 times;
5. Add 100µL prepared Detection Reagent B. Incubate 30 minutes at 37°C;
6. Aspirate and wash 5 times;
7. Add 90µL Substrate Solution. Incubate 10-20 minutes at 37°C;
8. Add 50µL Stop Solution. Read at 450nm immediately.

GIVEAWAYS

INCREMENT SERVICES

Magazine Citations
Mainz Identifizierung neuer potentieller Biomarker für das Kolonkarzinom sowie dessen Vorstufen Info:Source
Scientific Reports Target deletion of complement component 9 attenuates antibody-mediated hemolysis and lipopolysaccharide (LPS)-induced acute shock in mice pubmed:27444648
FASEB Journal Angiogenic factor imbalance precedes complement deposition in placentae of the BPH/5 model of preeclampsia Pubmed:29279353
Catalog No. Related products for research use of Mus musculus (Mouse) Organism species Applications (RESEARCH USE ONLY!)
RPB823Mu01 Recombinant Complement Component 9 (C9) Positive Control; Immunogen; SDS-PAGE; WB.
PAB823Mu01 Polyclonal Antibody to Complement Component 9 (C9) WB; IHC; ICC; IP.
SEB823Mu ELISA Kit for Complement Component 9 (C9) Enzyme-linked immunosorbent assay for Antigen Detection.
KSB823Mu01 ELISA Kit DIY Materials for Complement Component 9 (C9) Main materials for "Do It (ELISA Kit) Yourself".