ELISA Kit for Squamous Cell Carcinoma Antigen 1/2 (SCCA1/SCCA2)

SERPINB3; Serpin B3; SCCA1/SCCA2 fusion protein

Specificity

This assay has high sensitivity and excellent specificity for detection of Squamous Cell Carcinoma Antigen 1/2 (SCCA1/SCCA2).
No significant cross-reactivity or interference between Squamous Cell Carcinoma Antigen 1/2 (SCCA1/SCCA2) and analogues was observed.

Recovery

Matrices listed below were spiked with certain level of recombinant Squamous Cell Carcinoma Antigen 1/2 (SCCA1/SCCA2) and the recovery rates were calculated by comparing the measured value to the expected amount of Squamous Cell Carcinoma Antigen 1/2 (SCCA1/SCCA2) in samples.

Matrix Recovery range (%) Average(%)
serum(n=5) 80-93 83
EDTA plasma(n=5) 82-93 88
heparin plasma(n=5) 80-98 92

Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Squamous Cell Carcinoma Antigen 1/2 (SCCA1/SCCA2) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Squamous Cell Carcinoma Antigen 1/2 (SCCA1/SCCA2) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Squamous Cell Carcinoma Antigen 1/2 (SCCA1/SCCA2) and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Sample 1:2 1:4 1:8 1:16
serum(n=5) 85-98% 88-103% 83-104% 95-102%
EDTA plasma(n=5) 93-105% 87-101% 89-96% 86-101%
heparin plasma(n=5) 94-101% 83-98% 98-105% 90-97%

Stability

The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

Reagents and materials provided

Reagents Quantity Reagents Quantity
Pre-coated, ready to use 96-well strip plate 1 Plate sealer for 96 wells 4
Standard 2 Standard Diluent 1×20mL
Detection Reagent A 1×120µL Assay Diluent A 1×12mL
Detection Reagent B 1×120µL Assay Diluent B 1×12mL
TMB Substrate 1×9mL Stop Solution 1×6mL
Wash Buffer (30 × concentrate) 1×20mL Instruction manual 1

Assay procedure summary

1. Prepare all reagents, samples and standards;
2. Add 100µL standard or sample to each well. Incubate 1 hours at 37°C;
3. Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C;
4. Aspirate and wash 3 times;
5. Add 100µL prepared Detection Reagent B. Incubate 30 minutes at 37°C;
6. Aspirate and wash 5 times;
7. Add 90µL Substrate Solution. Incubate 10-20 minutes at 37°C;
8. Add 50µL Stop Solution. Read at 450nm immediately.

GIVEAWAYS

INCREMENT SERVICES

Catalog No. Related products for research use of Homo sapiens (Human) Organism species Applications (RESEARCH USE ONLY!)
RPB814Hu01 Recombinant Squamous Cell Carcinoma Antigen 1/2 (SCCA1/SCCA2) Positive Control; Immunogen; SDS-PAGE; WB.
RPB814Hu02 Recombinant Squamous Cell Carcinoma Antigen 1/2 (SCCA1/SCCA2) Positive Control; Immunogen; SDS-PAGE; WB.
PAB814Hu01 Polyclonal Antibody to Squamous Cell Carcinoma Antigen 1/2 (SCCA1/SCCA2) WB
PAB814Hu02 Polyclonal Antibody to Squamous Cell Carcinoma Antigen 1/2 (SCCA1/SCCA2) WB; IHC
LAB814Hu71 Biotin-Linked Polyclonal Antibody to Squamous Cell Carcinoma Antigen 1/2 (SCCA1/SCCA2) WB; IHC; ICC.
MAB814Hu23 Monoclonal Antibody to Squamous Cell Carcinoma Antigen 1/2 (SCCA1/SCCA2) WB; IHC; ICC; IP.
MAB814Hu24 Monoclonal Antibody to Squamous Cell Carcinoma Antigen 1/2 (SCCA1/SCCA2) WB; IHC; ICC; IP.
MAB814Hu21 Monoclonal Antibody to Squamous Cell Carcinoma Antigen 1/2 (SCCA1/SCCA2) WB; IHC; ICC; IP.
MAB814Hu22 Monoclonal Antibody to Squamous Cell Carcinoma Antigen 1/2 (SCCA1/SCCA2) WB.
MAB814Hu25 Monoclonal Antibody to Squamous Cell Carcinoma Antigen 1/2 (SCCA1/SCCA2) WB; IHC; ICC; IP.
SEB814Hu ELISA Kit for Squamous Cell Carcinoma Antigen 1/2 (SCCA1/SCCA2) Enzyme-linked immunosorbent assay for Antigen Detection.
LMB814Hu Multiplex Assay Kit for Squamous Cell Carcinoma Antigen 1/2 (SCCA1/SCCA2) ,etc. by FLIA (Flow Luminescence Immunoassay) FLIA Kit for Antigen Detection.
PSB814Hu01 Antibody Pair for Squamous Cell Carcinoma Antigen 1/2 (SCCA1/SCCA2) ELISA; CLIA; ELISPOT; Luminex; Immunochromatography and other Immunoassays.
KSB814Hu01 ELISA Kit DIY Materials for Squamous Cell Carcinoma Antigen 1/2 (SCCA1/SCCA2) Main materials for "Do It (ELISA Kit) Yourself".