ELISA Kit for Titin (TTN)

CMD1G; CMH9; CMPD4; HMERF; LGMD2J; TMD; Cardiomyopathy,Dilated 1G(Autosomal Dominant)

Specificity

This assay has high sensitivity and excellent specificity for detection of Titin (TTN).
No significant cross-reactivity or interference between Titin (TTN) and analogues was observed.

Recovery

Matrices listed below were spiked with certain level of recombinant Titin (TTN) and the recovery rates were calculated by comparing the measured value to the expected amount of Titin (TTN) in samples.

Matrix Recovery range (%) Average(%)
serum(n=5) 93-102 99
EDTA plasma(n=5) 87-99 95
heparin plasma(n=5) 78-101 82

Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Titin (TTN) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Titin (TTN) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Titin (TTN) and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Sample 1:2 1:4 1:8 1:16
serum(n=5) 86-98% 79-89% 85-97% 78-90%
EDTA plasma(n=5) 99-105% 93-102% 85-99% 89-103%
heparin plasma(n=5) 93-101% 97-105% 80-91% 87-105%

Stability

The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

Reagents and materials provided

Reagents Quantity Reagents Quantity
Pre-coated, ready to use 96-well strip plate 1 Plate sealer for 96 wells 4
Standard 2 Standard Diluent 1×20mL
Detection Reagent A 1×120µL Assay Diluent A 1×12mL
Detection Reagent B 1×120µL Assay Diluent B 1×12mL
TMB Substrate 1×9mL Stop Solution 1×6mL
Wash Buffer (30 × concentrate) 1×20mL Instruction manual 1

Assay procedure summary

1. Prepare all reagents, samples and standards;
2. Add 100µL standard or sample to each well. Incubate 1 hours at 37°C;
3. Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C;
4. Aspirate and wash 3 times;
5. Add 100µL prepared Detection Reagent B. Incubate 30 minutes at 37°C;
6. Aspirate and wash 5 times;
7. Add 90µL Substrate Solution. Incubate 10-20 minutes at 37°C;
8. Add 50µL Stop Solution. Read at 450nm immediately.

GIVEAWAYS

INCREMENT SERVICES

Magazine Citations
Proteome Science Proteomic analysis of effluents from perfused human heart for transplantation: identification of potential biomarkers for ischemic heart damag Proteomesci: 14775956
Catalog No. Related products for research use of Homo sapiens (Human) Organism species Applications (RESEARCH USE ONLY!)
RPB667Hu01 Recombinant Titin (TTN) Positive Control; Immunogen; SDS-PAGE; WB.
PAB667Hu01 Polyclonal Antibody to Titin (TTN) WB; IHC; ICC; IP.
MAB667Hu22 Monoclonal Antibody to Titin (TTN) WB; IHC; ICC; IP.
MAB667Hu21 Monoclonal Antibody to Titin (TTN) WB; IHC; ICC; IP.
LAB667Hu72 Biotin-Linked Monoclonal Antibody to Titin (TTN) WB; IHC; ICC.
SEB667Hu ELISA Kit for Titin (TTN) Enzyme-linked immunosorbent assay for Antigen Detection.
LMB667Hu Multiplex Assay Kit for Titin (TTN) ,etc. by FLIA (Flow Luminescence Immunoassay) FLIA Kit for Antigen Detection.
PSB667Hu01 Antibody Pair for Titin (TTN) ELISA; CLIA; ELISPOT; Luminex; Immunochromatography and other Immunoassays.
KSB667Hu01 ELISA Kit DIY Materials for Titin (TTN) Main materials for "Do It (ELISA Kit) Yourself".