Specificity

This assay has high sensitivity and excellent specificity for detection of Azurocidin (AZU).
No significant cross-reactivity or interference between Azurocidin (AZU) and analogues was observed.

Recovery

Matrices listed below were spiked with certain level of recombinant Azurocidin (AZU) and the recovery rates were calculated by comparing the measured value to the expected amount of Azurocidin (AZU) in samples.

Matrix Recovery range (%) Average(%)
serum(n=5) 95-105 101
EDTA plasma(n=5) 93-101 97
heparin plasma(n=5) 87-97 92

Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Azurocidin (AZU) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Azurocidin (AZU) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Azurocidin (AZU) and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Sample 1:2 1:4 1:8 1:16
serum(n=5) 85-97% 90-105% 85-92% 84-95%
EDTA plasma(n=5) 80-94% 86-97% 89-96% 80-90%
heparin plasma(n=5) 97-105% 78-93% 78-101% 82-103%

Stability

The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

Reagents and materials provided

Reagents Quantity Reagents Quantity
Pre-coated, ready to use 96-well strip plate 1 Plate sealer for 96 wells 4
Standard 2 Standard Diluent 1×20mL
Detection Reagent A 1×120µL Assay Diluent A 1×12mL
Detection Reagent B 1×120µL Assay Diluent B 1×12mL
TMB Substrate 1×9mL Stop Solution 1×6mL
Wash Buffer (30 × concentrate) 1×20mL Instruction manual 1

Assay procedure summary

1. Prepare all reagents, samples and standards;
2. Add 100µL standard or sample to each well. Incubate 1 hours at 37°C;
3. Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C;
4. Aspirate and wash 3 times;
5. Add 100µL prepared Detection Reagent B. Incubate 30 minutes at 37°C;
6. Aspirate and wash 5 times;
7. Add 90µL Substrate Solution. Incubate 10-20 minutes at 37°C;
8. Add 50µL Stop Solution. Read at 450nm immediately.

GIVEAWAYS

INCREMENT SERVICES

Magazine Citations
Journal of Maternal-Fetal and Neonatal Medicine Azurocidin levels in maternal serum in the first trimester can predict preterm prelabor rupture of membranes Pubmed: 23808364
Experimental and Therapeutic Medicine Azurocidin?induced inhibition of oxygen metabolism in mitochondria is antagonized by heparin Pubmed:25289044
Catalog No. Related products for research use of Homo sapiens (Human) Organism species Applications (RESEARCH USE ONLY!)
RPB461Hu01 Recombinant Azurocidin (AZU) Positive Control; Immunogen; SDS-PAGE; WB.
PAB461Hu01 Polyclonal Antibody to Azurocidin (AZU) WB,IHC
LAB461Hu71 Biotin-Linked Polyclonal Antibody to Azurocidin (AZU) WB; IHC; ICC.
MAB461Hu23 Monoclonal Antibody to Azurocidin (AZU) WB; IHC; ICC; IP.
MAB461Hu21 Monoclonal Antibody to Azurocidin (AZU) IHC
MAB461Hu27 Monoclonal Antibody to Azurocidin (AZU) WB; IHC; ICC; IP.
MAB461Hu22 Monoclonal Antibody to Azurocidin (AZU) WB
MAB461Hu24 Monoclonal Antibody to Azurocidin (AZU) WB.
MAB461Hu25 Monoclonal Antibody to Azurocidin (AZU) WB; IHC; ICC; IP.
FAB461Hu02 Anti-Azurocidin (AZU) Monoclonal Antibody FCM
FAB461Hu04 Anti-Azurocidin (AZU) Monoclonal Antibody FCM
SEB461Hu ELISA Kit for Azurocidin (AZU) Enzyme-linked immunosorbent assay for Antigen Detection.
LMB461Hu Multiplex Assay Kit for Azurocidin (AZU) ,etc. by FLIA (Flow Luminescence Immunoassay) FLIA Kit for Antigen Detection.
KSB461Hu01 ELISA Kit DIY Materials for Azurocidin (AZU) Main materials for "Do It(ELISA Kit) Yourself"