ELISA Kit for Mucin 4 (MUC4)

ASGP; Mucin 4 Cell Surface Associated; Mucin 4,Tracheobronchial; Testis mucin; Ascites sialoglycoprotein; Pancreatic adenocarcinoma mucin

Specificity

This assay has high sensitivity and excellent specificity for detection of Mucin 4 (MUC4).
No significant cross-reactivity or interference between Mucin 4 (MUC4) and analogues was observed.

Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Mucin 4 (MUC4) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Mucin 4 (MUC4) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%

Stability

The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

Reagents and materials provided

Reagents Quantity Reagents Quantity
Pre-coated, ready to use 96-well strip plate 1 Plate sealer for 96 wells 4
Standard 2 Standard Diluent 1×20mL
Detection Reagent A 1×120µL Assay Diluent A 1×12mL
Detection Reagent B 1×120µL Assay Diluent B 1×12mL
TMB Substrate 1×9mL Stop Solution 1×6mL
Wash Buffer (30 × concentrate) 1×20mL Instruction manual 1

Assay procedure summary

1. Prepare all reagents, samples and standards;
2. Add 100µL standard or sample to each well. Incubate 1 hours at 37°C;
3. Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C;
4. Aspirate and wash 3 times;
5. Add 100µL prepared Detection Reagent B. Incubate 30 minutes at 37°C;
6. Aspirate and wash 5 times;
7. Add 90µL Substrate Solution. Incubate 10-20 minutes at 37°C;
8. Add 50µL Stop Solution. Read at 450nm immediately.

GIVEAWAYS

INCREMENT SERVICES

Magazine Citations
Annals of surgical oncology Pancreatic cyst fluid and serum mucin levels predict dysplasia in intraductal papillary mucinous neoplasms of the pancreas. PubMed: 20717734
Exp Mol Pathol The regulatory role of SFRP5/WNT5A axis in allergic rhinitis through inhibiting JNK pathway activation and lowering mucin generation in human nasal?¡­ 33285209
ISCIENCE Identified human breastmilk compositions effectively inhibit SARS-CoV-2 and variants infection and replication Pubmed:35342878
Catalog No. Related products for research use of Mus musculus (Mouse) Organism species Applications (RESEARCH USE ONLY!)
RPB034Mu02 Recombinant Mucin 4 (MUC4) Positive Control; Immunogen; SDS-PAGE; WB.
RPB034Mu01 Recombinant Mucin 4 (MUC4) Positive Control; Immunogen; SDS-PAGE; WB.
PAB034Mu01 Polyclonal Antibody to Mucin 4 (MUC4) WB; IHC; ICC; IP.
PAB034Mu02 Polyclonal Antibody to Mucin 4 (MUC4) WB; IHC; ICC; IP.
SEB034Mu ELISA Kit for Mucin 4 (MUC4) Enzyme-linked immunosorbent assay for Antigen Detection.
LMB034Mu Multiplex Assay Kit for Mucin 4 (MUC4) ,etc. by FLIA (Flow Luminescence Immunoassay) FLIA Kit for Antigen Detection.