Packages (Simulation)
Reagent Preparation
Image (I)
Image (II)
Certificate
ELISA Kit for Heat Shock Protein 90 (HSP90)
HtpG; Hsp-90 NTD; N-Terminal Domain
- Product No.SEA863Hu
- Organism SpeciesHomo sapiens (Human) Same name, Different species.
- Sample Typeserum, plasma, tissue homogenates, cell lysates, cell culture supernates and other biological fluids
- Test MethodDouble-antibody Sandwich
- Assay Length3h
- Detection Range3.12-200ng/mL
- SensitivityThe minimum detectable dose of this kit is typically less than 1.09ng/mL.
- DownloadInstruction Manual
- UOM 48T96T 96T*5 96T*10 96T*100
- FOB
US$ 441
For more details, please contact local distributors! US$ 630 US$ 2835 US$ 5355 US$ 44100
Specificity
This assay has high sensitivity and excellent specificity for detection of Heat Shock Protein 90 (HSP90).
No significant cross-reactivity or interference between Heat Shock Protein 90 (HSP90) and analogues was observed.
Recovery
Matrices listed below were spiked with certain level of recombinant Heat Shock Protein 90 (HSP90) and the recovery rates were calculated by comparing the measured value to the expected amount of Heat Shock Protein 90 (HSP90) in samples.
Matrix | Recovery range (%) | Average(%) |
serum(n=5) | 81-94 | 89 |
EDTA plasma(n=5) | 80-103 | 95 |
heparin plasma(n=5) | 89-103 | 101 |
Precision
Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Heat Shock Protein 90 (HSP90) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Heat Shock Protein 90 (HSP90) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%
Linearity
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Heat Shock Protein 90 (HSP90) and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Sample | 1:2 | 1:4 | 1:8 | 1:16 |
serum(n=5) | 80-104% | 96-105% | 88-101% | 86-93% |
EDTA plasma(n=5) | 78-105% | 78-94% | 89-97% | 96-103% |
heparin plasma(n=5) | 96-105% | 87-105% | 78-104% | 79-99% |
Stability
The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.
Reagents and materials provided
Reagents | Quantity | Reagents | Quantity |
Pre-coated, ready to use 96-well strip plate | 1 | Plate sealer for 96 wells | 4 |
Standard | 2 | Standard Diluent | 1×20mL |
Detection Reagent A | 1×120µL | Assay Diluent A | 1×12mL |
Detection Reagent B | 1×120µL | Assay Diluent B | 1×12mL |
TMB Substrate | 1×9mL | Stop Solution | 1×6mL |
Wash Buffer (30 × concentrate) | 1×20mL | Instruction manual | 1 |
Assay procedure summary
1. Prepare all reagents, samples and standards;
2. Add 100µL standard or sample to each well. Incubate 1 hours at 37°C;
3. Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C;
4. Aspirate and wash 3 times;
5. Add 100µL prepared Detection Reagent B. Incubate 30 minutes at 37°C;
6. Aspirate and wash 5 times;
7. Add 90µL Substrate Solution. Incubate 10-20 minutes at 37°C;
8. Add 50µL Stop Solution. Read at 450nm immediately.
GIVEAWAYS
INCREMENT SERVICES
Magazine | Citations |
Arch Bronconeumol | Concentraciones de proteínas de estrés térmico, L-arginina y dimetilarginina asimétrica en pacientes con síndrome de apnea obstructiva del sue PubMed: 26055465 |
Arch Bronconeumol | Heat Shock Proteins, PubMed: 26055465 |
Jurnal Teknologi | Apigenin increases cisplatin inhibitory effects on the telomerase activity of triple negative breast cancer cells : |
博士论文 | PHYSIOLOGICAL ADAPTATIONS TO HEAT ACCLIMATION; REPERCUSSIONS ON CYCLING PERFORMANCE |
Catalog No. | Related products for research use of Homo sapiens (Human) Organism species | Applications (RESEARCH USE ONLY!) |
PAA863Hu01 | Polyclonal Antibody to Heat Shock Protein 90 (HSP90) | WB; IHC; ICC; IP. |
FAA863Hu51 | APC-Linked Anti-Heat Shock Protein 90 (HSP90) Polyclonal Antibody | Flow cytometry. |
HEA863Hu | High Sensitive ELISA Kit for Heat Shock Protein 90 (HSP90) | Enzyme-linked immunosorbent assay for Antigen Detection. |
SEA863Hu | ELISA Kit for Heat Shock Protein 90 (HSP90) | Enzyme-linked immunosorbent assay for Antigen Detection. |
LMA863Hu | Multiplex Assay Kit for Heat Shock Protein 90 (HSP90) ,etc. by FLIA (Flow Luminescence Immunoassay) | FLIA Kit for Antigen Detection. |