ELISA Kit for Coagulation Factor V (F5)

FV; FVL; PCCF; Proaccelerin; Labile Factor; Factor V Leiden; Accelerator Globulin; Activated protein C cofactor; Proaccelerin, labile factor

Specificity

This assay has high sensitivity and excellent specificity for detection of Coagulation Factor V (F5).
No significant cross-reactivity or interference between Coagulation Factor V (F5) and analogues was observed.

Recovery

Matrices listed below were spiked with certain level of recombinant Coagulation Factor V (F5) and the recovery rates were calculated by comparing the measured value to the expected amount of Coagulation Factor V (F5) in samples.

Matrix Recovery range (%) Average(%)
sodium citrate plasma(n=5) 80-91 84

Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Coagulation Factor V (F5) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Coagulation Factor V (F5) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Coagulation Factor V (F5) and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Sample 1:2 1:4 1:8 1:16
sodium citrate plasma(n=5) 80-89% 88-95% 83-90% 78-94%

Stability

The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

Reagents and materials provided

Reagents Quantity Reagents Quantity
Pre-coated, ready to use 96-well strip plate 1 Plate sealer for 96 wells 4
Standard 2 Standard Diluent 1×20mL
Detection Reagent A 1×120µL Assay Diluent A 1×12mL
Detection Reagent B 1×120µL Assay Diluent B 1×12mL
TMB Substrate 1×9mL Stop Solution 1×6mL
Wash Buffer (30 × concentrate) 1×20mL Instruction manual 1

Assay procedure summary

1. Prepare all reagents, samples and standards;
2. Add 100µL standard or sample to each well. Incubate 1 hours at 37°C;
3. Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C;
4. Aspirate and wash 3 times;
5. Add 100µL prepared Detection Reagent B. Incubate 30 minutes at 37°C;
6. Aspirate and wash 5 times;
7. Add 90µL Substrate Solution. Incubate 10-20 minutes at 37°C;
8. Add 50µL Stop Solution. Read at 450nm immediately.

GIVEAWAYS

INCREMENT SERVICES

Magazine Citations
Biochem Biophys Res Commun Thrombin generation in abdominal sepsis is Rho-kinase-dependent PubMed: 25813486
Cell Transplantation Evidence of a Pivotal Role for the Distal Part of the Complement Cascade in the Diurnal Release of Hematopoietic Stem Cells Into Peripheral Blood Pubmed:26087465
Cardiovascular Research Loss of accelerates arterial thrombosis by increasing formation of neutrophil extracellular traps and plasma tissue factor activity Pubmed:29444200
Catalog No. Related products for research use of Rattus norvegicus (Rat) Organism species Applications (RESEARCH USE ONLY!)
RPA675Ra01 Recombinant Coagulation Factor V (F5) Positive Control; Immunogen; SDS-PAGE; WB.
PAA675Ra01 Polyclonal Antibody to Coagulation Factor V (F5) WB; IHC; ICC; IP.
MAA675Ra21 Monoclonal Antibody to Coagulation Factor V (F5) WB; IHC; ICC; IP.
SEA675Ra ELISA Kit for Coagulation Factor V (F5) Enzyme-linked immunosorbent assay for Antigen Detection.
LMA675Ra Multiplex Assay Kit for Coagulation Factor V (F5) ,etc. by FLIA (Flow Luminescence Immunoassay) FLIA Kit for Antigen Detection.