ELISA Kit for Oncostatin M (OSM)

OS-M

Specificity

This assay has high sensitivity and excellent specificity for detection of Oncostatin M (OSM).
No significant cross-reactivity or interference between Oncostatin M (OSM) and analogues was observed.

Recovery

Matrices listed below were spiked with certain level of recombinant Oncostatin M (OSM) and the recovery rates were calculated by comparing the measured value to the expected amount of Oncostatin M (OSM) in samples.

Matrix Recovery range (%) Average(%)
serum(n=5) 84-99 89
EDTA plasma(n=5) 94-103 101
heparin plasma(n=5) 90-97 94

Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Oncostatin M (OSM) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Oncostatin M (OSM) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Oncostatin M (OSM) and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Sample 1:2 1:4 1:8 1:16
serum(n=5) 86-97% 81-97% 79-92% 90-97%
EDTA plasma(n=5) 80-99% 96-105% 94-102% 93-105%
heparin plasma(n=5) 88-104% 91-101% 79-94% 82-103%

Stability

The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

Reagents and materials provided

Reagents Quantity Reagents Quantity
Pre-coated, ready to use 96-well strip plate 1 Plate sealer for 96 wells 4
Standard 2 Standard Diluent 1×20mL
Detection Reagent A 1×120µL Assay Diluent A 1×12mL
Detection Reagent B 1×120µL Assay Diluent B 1×12mL
TMB Substrate 1×9mL Stop Solution 1×6mL
Wash Buffer (30 × concentrate) 1×20mL Instruction manual 1

Assay procedure summary

1. Prepare all reagents, samples and standards;
2. Add 100µL standard or sample to each well. Incubate 1 hours at 37°C;
3. Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C;
4. Aspirate and wash 3 times;
5. Add 100µL prepared Detection Reagent B. Incubate 30 minutes at 37°C;
6. Aspirate and wash 5 times;
7. Add 90µL Substrate Solution. Incubate 10-20 minutes at 37°C;
8. Add 50µL Stop Solution. Read at 450nm immediately.

GIVEAWAYS

INCREMENT SERVICES

Magazine Citations
Journal of Obstetrics and Gynaecology Research Oncostatin M as a target biological molecule of preeclampsia PubMed: 20149034
J Neuroinflammation Oncostatin M promotes excitotoxicity by inhibiting glutamate uptake in astrocytes: implications in HIV-associated neurotoxicity via Smad2/4 in Human Renal Proximal Tubular Epithelial Cells Pubmed:27287400
22 Modulating the phenotype of host macrophages to enhance osteogenesis in MSC-laden hydrogels: Design of a glucomannan coating material. pubmed:28582717
Advanced Functional Materials Fungal Component Coating Enhances Titanium Implant‐Bone Integration Doi: 10.1002/adfm.201804483
Catalog No. Related products for research use of Bos taurus; Bovine (Cattle) Organism species Applications (RESEARCH USE ONLY!)
RPA110Bo01 Recombinant Oncostatin M (OSM) Positive Control; Immunogen; SDS-PAGE; WB.
PAA110Bo01 Polyclonal Antibody to Oncostatin M (OSM) WB; IHC; ICC; IP.
SEA110Bo ELISA Kit for Oncostatin M (OSM) Enzyme-linked immunosorbent assay for Antigen Detection.
LMA110Bo Multiplex Assay Kit for Oncostatin M (OSM) ,etc. by FLIA (Flow Luminescence Immunoassay) FLIA Kit for Antigen Detection.