CLIA Kit for Interleukin 20 (IL20)

IL10D; IL10-D; ZCYTO10; Cytokine Zcyto10

Specificity

This assay has high sensitivity and excellent specificity for detection of Interleukin 20 (IL20).
No significant cross-reactivity or interference between Interleukin 20 (IL20) and analogues was observed.

Recovery

Matrices listed below were spiked with certain level of recombinant Interleukin 20 (IL20) and the recovery rates were calculated by comparing the measured value to the expected amount of Interleukin 20 (IL20) in samples.

Matrix Recovery range (%) Average(%)
serum(n=5) 95-102 98
EDTA plasma(n=5) 80-95 87
heparin plasma(n=5) 93-101 97

Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Interleukin 20 (IL20) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Interleukin 20 (IL20) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Interleukin 20 (IL20) and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Sample 1:2 1:4 1:8 1:16
serum(n=5) 81-96% 81-95% 96-105% 98-105%
EDTA plasma(n=5) 84-96% 83-97% 78-101% 81-101%
heparin plasma(n=5) 79-94% 88-95% 84-97% 81-97%

Stability

The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

Reagents and materials provided

Reagents Quantity Reagents Quantity
Pre-coated, ready to use 96-well strip plate 1 Plate sealer for 96 wells 4
Standard 2 Standard Diluent 1×20mL
Detection Reagent A 1×120µL Assay Diluent A 1×12mL
Detection Reagent B 1×120µL Assay Diluent B 1×12mL
Substrate A 1×10mL Substrate B 1×2mL
Wash Buffer (30 × concentrate) 1×20mL Instruction manual 1

Assay procedure summary

1. Prepare all reagents, samples and standards;
2. Add 100µL standard or sample to each well. Incubate 1 hours at 37°C;
3. Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C;
4. Aspirate and wash 3 times;
5. Add 100µL prepared Detection Reagent B. Incubate 30 minutes at 37°C;
6. Aspirate and wash 5 times;
7. Add 100µL Substrate Solution. Incubate 10 minutes at 37°C;
8. Read RLU value immediately.

GIVEAWAYS

INCREMENT SERVICES

Magazine Citations
Molecular Oncology The tumor promoting activity of the EP4 receptor for prostaglandin E2 in murine skin Pubmed:25034079
Biochimica et Biophysica Acta (BBA) - Molecular Basis of Disease IL-20 promotes hypoxia/reoxygenation-induced mitochondrial dysfunction and apoptosis in cardiomyocytes by upregulating oxidative stress by activating the PKC …
Theranostics IL20RA signaling enhances stemness and promotes the formation of an immunosuppressive microenvironment in breast cancer 33456560
Catalog No. Related products for research use of Homo sapiens (Human) Organism species Applications (RESEARCH USE ONLY!)
RPC058Hu01 Recombinant Interleukin 20 (IL20) Positive Control; Immunogen; SDS-PAGE; WB.
APC058Hu01 Active Interleukin 20 (IL20) Cell culture; Activity Assays.
PAC058Hu01 Polyclonal Antibody to Interleukin 20 (IL20) WB; IHC; ICC; IP.
MAC058Hu22 Monoclonal Antibody to Interleukin 20 (IL20) WB; IHC; ICC; IP.
RAC058Hu21 Recombinant Antibody to Interleukin 20 (IL20) WB; IF; ICC; IHC; IP; FCM.
SEC058Hu ELISA Kit for Interleukin 20 (IL20) Enzyme-linked immunosorbent assay for Antigen Detection.
SCC058Hu CLIA Kit for Interleukin 20 (IL20) Chemiluminescent immunoassay for Antigen Detection.
LMC058Hu Multiplex Assay Kit for Interleukin 20 (IL20) ,etc. by FLIA (Flow Luminescence Immunoassay) FLIA Kit for Antigen Detection.