OneStep ELISA Kit for Chemokine (C-C motif) ligand 6 (CCL6)

MRP1; SCYA6; C10; Macrophage Inflammatory Protein Related Protein 1; Small Inducible Cytokine A6

Specificity

This assay has high sensitivity and excellent specificity for detection of Chemokine (C-C motif) ligand 6 (CCL6).
No significant cross-reactivity or interference between Chemokine (C-C motif) ligand 6 (CCL6) and analogues was observed.

Recovery

Matrices listed below were spiked with certain level of recombinant Chemokine (C-C motif) ligand 6 (CCL6) and the recovery rates were calculated by comparing the measured value to the expected amount of Chemokine (C-C motif) ligand 6 (CCL6) in samples.

Matrix Recovery range (%) Average(%)
serum(n=5) 80-91 87
EDTA plasma(n=5) 83-96 90
heparin plasma(n=5) 78-92 88

Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Chemokine (C-C motif) ligand 6 (CCL6) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Chemokine (C-C motif) ligand 6 (CCL6) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Chemokine (C-C motif) ligand 6 (CCL6) and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Sample 1:2 1:4 1:8 1:16
serum(n=5) 79-104% 79-95% 80-101% 99-105%
EDTA plasma(n=5) 96-103% 81-95% 91-99% 93-101%
heparin plasma(n=5) 91-102% 93-105% 81-90% 81-95%

Stability

The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

Reagents and materials provided

Reagents Quantity Reagents Quantity

Assay procedure summary

1. Prepare all reagents, samples and standards;
2. Add 50µL standard or sample to each well, then add 50µL prepared Detection Reagent A to each well. Incubate 1 hour at 37°C;
3. Aspirate and wash 3 times;
4. Add 90µL Substrate Solution. Incubate 10-20 minutes at 37°C;
5. Add 50µL Stop Solution. Read at 450nm immediately.

GIVEAWAYS

INCREMENT SERVICES

Catalog No. Related products for research use of Mus musculus (Mouse) Organism species Applications (RESEARCH USE ONLY!)
RPB060Mu01 Recombinant Chemokine (C-C motif) ligand 6 (CCL6) Positive Control; Immunogen; SDS-PAGE; WB.
PAB060Mu01 Polyclonal Antibody to Chemokine (C-C motif) ligand 6 (CCL6) WB
OEB060Mu OneStep ELISA Kit for Chemokine (C-C motif) ligand 6 (CCL6) Enzyme-linked immunosorbent assay for Antigen Detection.
SEB060Mu ELISA Kit for Chemokine (C-C motif) ligand 6 (CCL6) Enzyme-linked immunosorbent assay for Antigen Detection.
MEB060Mu Mini Samples ELISA Kit for Chemokine (C-C motif) ligand 6 (CCL6) Enzyme-linked immunosorbent assay for Antigen Detection.
CMB060Mu Multiplex Assay Kit for Chemokine (C-C motif) ligand 6 (CCL6) ,etc. by CBA (Cytometric Bead Array) FLIA Kit for Antigen Detection.
LMB060Mu Multiplex Assay Kit for Chemokine (C-C motif) ligand 6 (CCL6) ,etc. by FLIA (Flow Luminescence Immunoassay) FLIA Kit for Antigen Detection.