Mini Samples ELISA Kit for Coagulation Factor X (F10)

FX; FXA; Stuart Prower Factor; Thrombokinase

Specificity

This assay has high sensitivity and excellent specificity for detection of Mini Samples Coagulation Factor X (F10).
No significant cross-reactivity or interference between Mini Samples Coagulation Factor X (F10) and analogues was observed.

Recovery

Matrices listed below were spiked with certain level of recombinant Mini Samples Coagulation Factor X (F10) and the recovery rates were calculated by comparing the measured value to the expected amount of Mini Samples Coagulation Factor X (F10) in samples.

Matrix Recovery range (%) Average(%)
sodium citrate plasma(n=5) 98-105 102

Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Mini Samples Coagulation Factor X (F10) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Mini Samples Coagulation Factor X (F10) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Mini Samples Coagulation Factor X (F10) and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Sample 1:2 1:4 1:8 1:16
sodium citrate plasma(n=5) 78-98% 85-101% 78-105% 79-93%

Stability

The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

Reagents and materials provided

Reagents Quantity Reagents Quantity
Pre-coated, ready to use 96-well strip plate 1 Plate sealer for 96 wells 4
Standard 2 Standard Diluent 1×20mL
Detection Reagent A 1×60µL Assay Diluent A 1×6mL
Detection Reagent B 1×60µL Assay Diluent B 1×6mL
TMB Substrate 1×4.5mL Stop Solution 1×3mL
Wash Buffer (30 × concentrate) 1×10mL Instruction manual 1

Assay procedure summary

1. Prepare all reagents, samples and standards;
2. Add 25µL standard or sample to each well. Incubate 1 hour at 37°C;
3. Aspirate and add 25µL prepared Detection Reagent A. Incubate 1 hour at 37°C;
4. Aspirate and wash 3 times;
5. Add 25µL prepared Detection Reagent B. Incubate 30 minutes at 37°C;
6. Aspirate and wash 5 times;
7. Add 25µL Substrate Solution. Incubate 10-20 minutes at 37°C;
8. Add 20µL Stop Solution. Read at 450nm immediately.

GIVEAWAYS

INCREMENT SERVICES

Magazine Citations
Am J Physiol Heart Circ Physiol. Monocytes regulate systemic coagulation and inflammation in abdominal sepsis Pubmed:25502108
Biochem Biophys Res Commun Thrombin generation in abdominal sepsis is Rho-kinase-dependent PubMed: 25813486
Catalog No. Related products for research use of Mus musculus (Mouse) Organism species Applications (RESEARCH USE ONLY!)
RPA834Mu01 Recombinant Coagulation Factor X (F10) Positive Control; Immunogen; SDS-PAGE; WB.
RPA834Mu02 Recombinant Coagulation Factor X (F10) Positive Control; Immunogen; SDS-PAGE; WB.
PAA834Mu01 Polyclonal Antibody to Coagulation Factor X (F10) WB; IHC; ICC; IP.
PAA834Mu02 Polyclonal Antibody to Coagulation Factor X (F10) WB; IHC; ICC; IP.
MEA834Mu Mini Samples ELISA Kit for Coagulation Factor X (F10) Enzyme-linked immunosorbent assay for Antigen Detection.
SEA834Mu ELISA Kit for Coagulation Factor X (F10) Enzyme-linked immunosorbent assay for Antigen Detection.
LMA834Mu Multiplex Assay Kit for Coagulation Factor X (F10) ,etc. by FLIA (Flow Luminescence Immunoassay) FLIA Kit for Antigen Detection.
KSA834Mu01 ELISA Kit DIY Materials for Coagulation Factor X (F10) Main materials for "Do It (ELISA Kit) Yourself".