Multiplex Assay Kit for Lysophosphatidic Acid (LPA) ,etc. by FLIA (Flow Luminescence Immunoassay)

LysoPA; 1-Oleoyl-sn-Glycero-3-Phosphate; Lysophosphatidyl Acid

(Note: Up to 8-plex in one testing reaction)

Specificity

This assay has high sensitivity and excellent specificity for detection of Lysophosphatidic Acid (LPA) ,etc. by FLIA (Flow Luminescence Immunoassay).
No significant cross-reactivity or interference between Lysophosphatidic Acid (LPA) ,etc. by FLIA (Flow Luminescence Immunoassay) and analogues was observed.

Recovery

Matrices listed below were spiked with certain level of Lysophosphatidic Acid (LPA) ,etc. by FLIA (Flow Luminescence Immunoassay) and the recovery rates were calculated by comparing the measured value to the expected amount of Lysophosphatidic Acid (LPA) ,etc. by FLIA (Flow Luminescence Immunoassay) in samples.

Matrix Recovery range (%) Average(%)
serum(n=5) 94-104 99
EDTA plasma(n=5) 98-105 102
heparin plasma(n=5) 81-90 86

Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Lysophosphatidic Acid (LPA) ,etc. by FLIA (Flow Luminescence Immunoassay) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Lysophosphatidic Acid (LPA) ,etc. by FLIA (Flow Luminescence Immunoassay) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Lysophosphatidic Acid (LPA) ,etc. by FLIA (Flow Luminescence Immunoassay) and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Sample 1:2 1:4 1:8 1:16
serum(n=5) 88-101% 78-103% 79-93% 83-103%
EDTA plasma(n=5) 91-101% 97-105% 90-97% 85-94%
heparin plasma(n=5) 91-99% 82-97% 81-104% 96-104%

Stability

The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

Reagents and materials provided

Reagents Quantity Reagents Quantity
96-well plate 1 Plate sealer for 96 wells 4
Pre-Mixed Standard 2 Standard Diluent 1×20mL
Pre-Mixed Magnetic beads (22#:LPA) 1 Analysis buffer 1×20mL
Pre-Mixed Detection Reagent A 1×120μL Assay Diluent A 1×12mL
Detection Reagent B (PE-SA) 1×120μL Assay Diluent B 1×12mL
Sheath Fluid 1×10mL Wash Buffer (30 × concentrate) 1×20mL
Instruction manual 1

Assay procedure summary

1. Preparation of standards, reagents and samples before the experiment;
2. Add 50μL standard or sample to each well,
    add 10μL magnetic beads,and 50μL Detection Reagent A,incubate 60min at 37°C on shaker;
3. Wash plate on magnetic frame for three times;
4. Add 100μL prepared Detection Reagent B, and incubate 30 min at 37°C on shaker;
5. Wash plate on magnetic frame for three times;
6. Add 100μL sheath solution, swirl for 2 minutes, read on the machine.

GIVEAWAYS

INCREMENT SERVICES

Magazine Citations
Molecular Sciences Pigment Epithelium-Derived Factor (PEDF) Protects Osteoblastic Cell Line from Glucocorticoid-Induced Apoptosis via PEDF-R Pubmed:27187377
Nature Communications ER-residential Nogo-B accelerates NAFLD-associated HCC mediated by metabolic reprogramming of oxLDL lipophagy Pubmed: 31358770
JOURNAL OF THROMBOSIS AND HAEMOSTASIS Lysophosphatidic acid promotes thrombus stability by inducing rapid formation of neutrophil extracellular traps: A new mechanism of thrombosis Pubmed: 32291893
Signal Transduction and Targeted Therapy The Agpat4/LPA axis in colorectal cancer cells regulates antitumor responses via p38/p65 signaling in macrophages Pubmed: 32296017
Journal of Neuroinflammation MAPK signaling determines lysophosphatidic acid (LPA)-induced inflammation in microglia Pubmed: 32326963
Catalog No. Related products for research use of Pan-species (General) Organism species Applications (RESEARCH USE ONLY!)
CEK623Ge ELISA Kit for Lysophosphatidic Acid (LPA) Enzyme-linked immunosorbent assay for Antigen Detection.
LMK623Ge Multiplex Assay Kit for Lysophosphatidic Acid (LPA) ,etc. by FLIA (Flow Luminescence Immunoassay) FLIA Kit for Antigen Detection.