Multiplex Assay Kit for Suprabasin (SBSN) ,etc. by FLIA (Flow Luminescence Immunoassay)

(Note: Up to 8-plex in one testing reaction)

Specificity

This assay has high sensitivity and excellent specificity for detection of Suprabasin (SBSN) ,etc. by FLIA (Flow Luminescence Immunoassay).
No significant cross-reactivity or interference between Suprabasin (SBSN) ,etc. by FLIA (Flow Luminescence Immunoassay) and analogues was observed.

Recovery

Matrices listed below were spiked with certain level of recombinant Suprabasin (SBSN) ,etc. by FLIA (Flow Luminescence Immunoassay) and the recovery rates were calculated by comparing the measured value to the expected amount of Suprabasin (SBSN) ,etc. by FLIA (Flow Luminescence Immunoassay) in samples.

Matrix Recovery range (%) Average(%)
serum(n=5) 80-105 88
EDTA plasma(n=5) 85-99 90
heparin plasma(n=5) 89-103 99

Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Suprabasin (SBSN) ,etc. by FLIA (Flow Luminescence Immunoassay) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Suprabasin (SBSN) ,etc. by FLIA (Flow Luminescence Immunoassay) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Suprabasin (SBSN) ,etc. by FLIA (Flow Luminescence Immunoassay) and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Sample 1:2 1:4 1:8 1:16
serum(n=5) 78-94% 88-101% 92-105% 83-103%
EDTA plasma(n=5) 93-102% 99-105% 92-99% 89-103%
heparin plasma(n=5) 99-105% 97-105% 97-105% 81-93%

Stability

The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

Reagents and materials provided

Reagents Quantity Reagents Quantity
96-well plate 1 Plate sealer for 96 wells 4
Pre-Mixed Standard 2 Standard Diluent 1×20mL
Pre-Mixed Magnetic beads (22#:SBSN) 1 Analysis buffer 1×20mL
Pre-Mixed Detection Reagent A 1×120μL Assay Diluent A 1×12mL
Detection Reagent B (PE-SA) 1×120μL Assay Diluent B 1×12mL
Sheath Fluid 1×10mL Wash Buffer (30 × concentrate) 1×20mL
Instruction manual 1

Assay procedure summary

1. Preparation of standards, reagents and samples before the experiment;
2. Add 100μL standard or sample to each well,
    add 10μL magnetic beads, and incubate 90min at 37°C on shaker;
3. Remove liquid on magnetic frame, add 100μL prepared Detection Reagent A. Incubate 60min at 37°C on shaker;
4. Wash plate on magnetic frame for three times;
5. Add 100μL prepared Detection Reagent B, and incubate 30 min at 37°C on shaker;
6. Wash plate on magnetic frame for three times;
7. Add 100μL sheath solution, swirl for 2 minutes, read on the machine.

GIVEAWAYS

INCREMENT SERVICES

Magazine Citations
Journal of Dermatological Science Decreased expression of suprabasin induces aberrant differentiation and apoptosis of epidermal keratinocytes: Possible role for atopic dermatitis Pubmed: 31399284
Catalog No. Related products for research use of Homo sapiens (Human) Organism species Applications (RESEARCH USE ONLY!)
RPH197Hu01 Recombinant Suprabasin (SBSN) Positive Control; Immunogen; SDS-PAGE; WB.
PAH197Hu01 Polyclonal Antibody to Suprabasin (SBSN) WB
MAH197Hu21 Monoclonal Antibody to Suprabasin (SBSN) WB; IHC; ICC; IP.
SEH197Hu ELISA Kit for Suprabasin (SBSN) Enzyme-linked immunosorbent assay for Antigen Detection.
LMH197Hu Multiplex Assay Kit for Suprabasin (SBSN) ,etc. by FLIA (Flow Luminescence Immunoassay) FLIA Kit for Antigen Detection.