Multiplex Assay Kit for S-Adenosyl Methionine (SAM) ,etc. by FLIA (Flow Luminescence Immunoassay)

SAMe; SAM-e; S-Adenosylmethionine; S-(5'-Adenosyl)-L-Methionine Chloride

(Note: Up to 8-plex in one testing reaction)

Specificity

This assay has high sensitivity and excellent specificity for detection of S-Adenosyl Methionine (SAM) ,etc. by FLIA (Flow Luminescence Immunoassay).
No significant cross-reactivity or interference between S-Adenosyl Methionine (SAM) ,etc. by FLIA (Flow Luminescence Immunoassay) and analogues was observed.

Recovery

Matrices listed below were spiked with certain level of S-Adenosyl Methionine (SAM) ,etc. by FLIA (Flow Luminescence Immunoassay) and the recovery rates were calculated by comparing the measured value to the expected amount of S-Adenosyl Methionine (SAM) ,etc. by FLIA (Flow Luminescence Immunoassay) in samples.

Matrix Recovery range (%) Average(%)
serum(n=5) 78-103 88
EDTA plasma(n=5) 88-95 91
heparin plasma(n=5) 92-101 96
sodium citrate plasma(n=5) 81-98 88

Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level S-Adenosyl Methionine (SAM) ,etc. by FLIA (Flow Luminescence Immunoassay) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level S-Adenosyl Methionine (SAM) ,etc. by FLIA (Flow Luminescence Immunoassay) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of S-Adenosyl Methionine (SAM) ,etc. by FLIA (Flow Luminescence Immunoassay) and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Sample 1:2 1:4 1:8 1:16
serum(n=5) 94-102% 96-104% 96-105% 79-101%
EDTA plasma(n=5) 86-97% 97-105% 79-93% 78-95%
heparin plasma(n=5) 90-103% 86-94% 92-102% 85-96%
sodium citrate plasma(n=5) 87-98% 92-99% 84-94% 97-104%

Stability

The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

Reagents and materials provided

Reagents Quantity Reagents Quantity
96-well plate 1 Plate sealer for 96 wells 4
Pre-Mixed Standard 2 Standard Diluent 1×20mL
Pre-Mixed Magnetic beads (22#:SAM) 1 Analysis buffer 1×20mL
Pre-Mixed Detection Reagent A 1×120μL Assay Diluent A 1×12mL
Detection Reagent B (PE-SA) 1×120μL Assay Diluent B 1×12mL
Sheath Fluid 1×10mL Wash Buffer (30 × concentrate) 1×20mL
Instruction manual 1

Assay procedure summary

1. Preparation of standards, reagents and samples before the experiment;
2. Add 50μL standard or sample to each well,
    add 10μL magnetic beads,and 50μL Detection Reagent A,incubate 60min at 37°C on shaker;
3. Wash plate on magnetic frame for three times;
4. Add 100μL prepared Detection Reagent B, and incubate 30 min at 37°C on shaker;
5. Wash plate on magnetic frame for three times;
6. Add 100μL sheath solution, swirl for 2 minutes, read on the machine.

GIVEAWAYS

INCREMENT SERVICES

Magazine Citations
Cell Death and Disease Hyperhomocysteinemia causes ER stress and impaired autophagy that is reversed by Vitamin B supplementation pubmed:27929536
PLoS One Attenuated expression of MTR in both prenatally androgenized mice and women with the hyperandrogenic phenotype of PCOS pubmed:29232372vvvv
Microbial Cell Factories Metabolic engineering of Acremonium chrysogenum for improving cephalosporin C production independent of methionine stimulation Pubmed:29879990
Cell Metabolism Serine metabolism antagonizes antiviral innate immunity by preventing ATP6V0d2-mediated YAP lysosomal degradation 33798471
Catalog No. Related products for research use of Pan-species (General) Organism species Applications (RESEARCH USE ONLY!)
CPG414Ge11 BSA Conjugated S-Adenosyl Methionine (SAM) Immunogen; SDS-PAGE; WB.
CPG414Ge21 OVA Conjugated S-Adenosyl Methionine (SAM) Immunogen; SDS-PAGE; WB.
PAG414Ge01 Polyclonal Antibody to S-Adenosyl Methionine (SAM) ELISA, CLIA. / IHC-Fr, ICC, IP (predicted).
PAG414Ge02 Polyclonal Antibody to S-Adenosyl Methionine (SAM) ELISA, CLIA. / IHC-Fr, ICC, IP (predicted).
MEG414Ge Mini Samples ELISA Kit for S-Adenosyl Methionine (SAM) Enzyme-linked immunosorbent assay for Antigen Detection.
CEG414Ge ELISA Kit for S-Adenosyl Methionine (SAM) Enzyme-linked immunosorbent assay for Antigen Detection.
LMG414Ge Multiplex Assay Kit for S-Adenosyl Methionine (SAM) ,etc. by FLIA (Flow Luminescence Immunoassay) FLIA Kit for Antigen Detection.
KSG414Ge11 ELISA Kit DIY Materials for S-Adenosyl Methionine (SAM) Main materials for "Do It (ELISA Kit) Yourself".