Multiplex Assay Kit for Immunoglobulin Kappa (Igk) ,etc. by FLIA (Flow Luminescence Immunoassay)

IGKC; HCAK1; Km; Immunoglobulin Kappa Constant; IgA Kappa

(Note: Up to 8-plex in one testing reaction)

Specificity

This assay has high sensitivity and excellent specificity for detection of Immunoglobulin Kappa (Igk).
No significant cross-reactivity or interference between Immunoglobulin Kappa (Igk) and analogues was observed.

Recovery

Matrices listed below were spiked with certain level of recombinant Immunoglobulin Kappa (Igk) and the recovery rates were calculated by comparing the measured value to the expected amount of Immunoglobulin Kappa (Igk) in samples.

Matrix Recovery range (%) Average(%)
serum(n=5) 97-105 101
EDTA plasma(n=5) 89-98 95
heparin plasma(n=5) 86-101 97
sodium citrate plasma(n=5) 81-91 88

Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Immunoglobulin Kappa (Igk) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Immunoglobulin Kappa (Igk) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Immunoglobulin Kappa (Igk) and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Sample 1:2 1:4 1:8 1:16
serum(n=5) 96-104% 80-103% 93-105% 96-105%
EDTA plasma(n=5) 87-94% 94-102% 86-95% 88-96%
heparin plasma(n=5) 99-105% 83-92% 92-99% 98-105%
sodium citrate plasma(n=5) 82-89% 91-105% 78-103% 96-104%

Stability

The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

Reagents and materials provided

Reagents Quantity Reagents Quantity
96-well plate 1 Plate sealer for 96 wells 4
Pre-Mixed Standard 2 Standard Diluent 1×20mL
Pre-Mixed Magnetic beads (22#:Igk) 1 Analysis buffer 1×20mL
Pre-Mixed Detection Reagent A 1×120μL Assay Diluent A 1×12mL
Detection Reagent B (PE-SA) 1×120μL Assay Diluent B 1×12mL
Sheath Fluid 1×10mL Wash Buffer (30 × concentrate) 1×20mL
Instruction manual 1

Assay procedure summary

1. Add 200μL analysis buffer solution to each well of the plate for pre-wetting;
2. Preparation of standards, reagents and samples before the experiment;
3. Add 100μL standard or sample to each well, add 10μL magnetic beads, and incubate 120 minutes protect from light at 25°C on shaker;
4. Remove liquid on magnetic frame, add 100μL prepared Detection Reagent A. Incubate 60 minutes protect from light at 25°C on shaker;
5. Remove liquid on magnetic frame, add 100μL prepared Detection Reagent B, and incubate 30 minutes protect from light at 25°C on shaker;
6. Wash plate once on magnetic frame;
7. Add 100μL sheath fluid, swirl for 10 minutes, read on the machine.

GIVEAWAYS

INCREMENT SERVICES

Magazine Citations
Cancer Cell Int Discovery and validation of FBLN1 and ANT3 as potential biomarkers for early detection of cervical cancer 33602229
Catalog No. Related products for research use of Homo sapiens (Human) Organism species Applications (RESEARCH USE ONLY!)
EPD038Hu61 Eukaryotic Immunoglobulin Kappa (Igk) Positive Control; Immunogen; SDS-PAGE; WB.
CPD038Hu21 OVA Conjugated Immunoglobulin Kappa (Igk) Immunogen; SDS-PAGE; WB.
PAD038Hu08 Polyclonal Antibody to Immunoglobulin Kappa (Igk) Testing in progress.
PAD038Hu06 Polyclonal Antibody to Immunoglobulin Kappa (Igk) WB
PAD038Hu01 Polyclonal Antibody to Immunoglobulin Kappa (Igk) WB; IHC; ICC; IP.
FAD038Hu01 Anti-Immunoglobulin Kappa (Igk) Polyclonal Antibody Flow cytometry.
SED038Hu ELISA Kit for Immunoglobulin Kappa (Igk) Enzyme-linked immunosorbent assay for Antigen Detection.