Multiplex Assay Kit for Fibrinogen Like Protein 1 (FGL1) ,etc. by FLIA (Flow Luminescence Immunoassay)

FGL-1; HFREP1; HP-041; LFIRE1; Hepassocin; Hepatocyte-derived fibrinogen-related protein 1; Liver fibrinogen-related protein 1

(Note: Up to 8-plex in one testing reaction)

Specificity

This assay has high sensitivity and excellent specificity for detection of Fibrinogen Like Protein 1 (FGL1) ,etc. by FLIA (Flow Luminescence Immunoassay).
No significant cross-reactivity or interference between Fibrinogen Like Protein 1 (FGL1) ,etc. by FLIA (Flow Luminescence Immunoassay) and analogues was observed.

Recovery

Matrices listed below were spiked with certain level of recombinant Fibrinogen Like Protein 1 (FGL1) ,etc. by FLIA (Flow Luminescence Immunoassay) and the recovery rates were calculated by comparing the measured value to the expected amount of Fibrinogen Like Protein 1 (FGL1) ,etc. by FLIA (Flow Luminescence Immunoassay) in samples.

Matrix Recovery range (%) Average(%)
serum(n=5) 78-95 82
EDTA plasma(n=5) 78-96 86
heparin plasma(n=5) 90-104 97

Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Fibrinogen Like Protein 1 (FGL1) ,etc. by FLIA (Flow Luminescence Immunoassay) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Fibrinogen Like Protein 1 (FGL1) ,etc. by FLIA (Flow Luminescence Immunoassay) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Fibrinogen Like Protein 1 (FGL1) ,etc. by FLIA (Flow Luminescence Immunoassay) and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Sample 1:2 1:4 1:8 1:16
serum(n=5) 86-105% 86-94% 86-103% 78-102%
EDTA plasma(n=5) 86-94% 78-97% 90-103% 94-102%
heparin plasma(n=5) 80-94% 95-105% 81-93% 96-103%

Stability

The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

Reagents and materials provided

Reagents Quantity Reagents Quantity
96-well plate 1 Plate sealer for 96 wells 4
Pre-Mixed Standard 2 Standard Diluent 1×20mL
Pre-Mixed Magnetic beads (22#:FGL1) 1 Analysis buffer 1×20mL
Pre-Mixed Detection Reagent A 1×120μL Assay Diluent A 1×12mL
Detection Reagent B (PE-SA) 1×120μL Assay Diluent B 1×12mL
Sheath Fluid 1×10mL Wash Buffer (30 × concentrate) 1×20mL
Instruction manual 1

Assay procedure summary

1. Preparation of standards, reagents and samples before the experiment;
2. Add 100μL standard or sample to each well,
    add 10μL magnetic beads, and incubate 90min at 37°C on shaker;
3. Remove liquid on magnetic frame, add 100μL prepared Detection Reagent A. Incubate 60min at 37°C on shaker;
4. Wash plate on magnetic frame for three times;
5. Add 100μL prepared Detection Reagent B, and incubate 30 min at 37°C on shaker;
6. Wash plate on magnetic frame for three times;
7. Add 100μL sheath solution, swirl for 2 minutes, read on the machine.

GIVEAWAYS

INCREMENT SERVICES

Magazine Citations
J Hepatol. The role of Hepassocin in the development of non-alcoholic fatty liver disease Pubmed: 23792031
Diabetes Management Serum hepassocin concentrations in diabetic patients with or without nonalcoholic fatty liver disease Futuremedicine:Source
United States Patent Application Compositions and Methods for Diagnosing Lung Cancer y2016:0077095.html
Cell Obesity Drives STAT-1-Dependent NASH and STAT-3-Dependent HCC Doi: 10.1016/j.cell.2018.09.053
Liver Research Do different bariatric surgery procedures impact hepassocin plasma levels in patients with type 2 diabetes mellitus?
Impacts of Different Modes of Bariatric Surgery on Plasma Levels of Hepassocin in Patients with Diabetes Mellitus
Catalog No. Related products for research use of Rattus norvegicus (Rat) Organism species Applications (RESEARCH USE ONLY!)
RPD022Ra01 Recombinant Fibrinogen Like Protein 1 (FGL1) Positive Control; Immunogen; SDS-PAGE; WB.
RPD022Ra02 Recombinant Fibrinogen Like Protein 1 (FGL1) Positive Control; Immunogen; SDS-PAGE; WB.
PAD022Ra01 Polyclonal Antibody to Fibrinogen Like Protein 1 (FGL1) WB; IHC; ICC; IP.
SED022Ra ELISA Kit for Fibrinogen Like Protein 1 (FGL1) Enzyme-linked immunosorbent assay for Antigen Detection.
LMD022Ra Multiplex Assay Kit for Fibrinogen Like Protein 1 (FGL1) ,etc. by FLIA (Flow Luminescence Immunoassay) FLIA Kit for Antigen Detection.
KSD022Ra01 ELISA Kit DIY Materials for Fibrinogen Like Protein 1 (FGL1) Main materials for "Do It (ELISA Kit) Yourself".