Multiplex Assay Kit for Xanthine Dehydrogenase (XDH) ,etc. by FLIA (Flow Luminescence Immunoassay)

XO; XOR; XOD; Xanthine Oxidase; NAD+-Xanthine Dehydrogenase; Xanthine-NAD+ Oxidoreductase; Xanthine/NAD+ Oxidoreductase; Xanthine Oxidoreductase

(Note: Up to 8-plex in one testing reaction)

Specificity

This assay has high sensitivity and excellent specificity for detection of Xanthine Dehydrogenase (XDH) ,etc. by FLIA (Flow Luminescence Immunoassay).
No significant cross-reactivity or interference between Xanthine Dehydrogenase (XDH) ,etc. by FLIA (Flow Luminescence Immunoassay) and analogues was observed.

Recovery

Matrices listed below were spiked with certain level of recombinant Xanthine Dehydrogenase (XDH) ,etc. by FLIA (Flow Luminescence Immunoassay) and the recovery rates were calculated by comparing the measured value to the expected amount of Xanthine Dehydrogenase (XDH) ,etc. by FLIA (Flow Luminescence Immunoassay) in samples.

Matrix Recovery range (%) Average(%)
serum(n=5) 97-105 101
EDTA plasma(n=5) 96-105 102
heparin plasma(n=5) 81-95 89
sodium citrate plasma(n=5) 84-99 90

Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Xanthine Dehydrogenase (XDH) ,etc. by FLIA (Flow Luminescence Immunoassay) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Xanthine Dehydrogenase (XDH) ,etc. by FLIA (Flow Luminescence Immunoassay) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Xanthine Dehydrogenase (XDH) ,etc. by FLIA (Flow Luminescence Immunoassay) and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Sample 1:2 1:4 1:8 1:16
serum(n=5) 83-95% 93-102% 88-101% 85-92%
EDTA plasma(n=5) 80-102% 79-93% 95-102% 84-99%
heparin plasma(n=5) 98-105% 82-95% 80-96% 98-105%
sodium citrate plasma(n=5) 84-104% 82-89% 78-104% 98-105%

Stability

The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

Reagents and materials provided

Reagents Quantity Reagents Quantity
96-well plate 1 Plate sealer for 96 wells 4
Pre-Mixed Standard 2 Standard Diluent 1×20mL
Pre-Mixed Magnetic beads (22#:XDH) 1 Analysis buffer 1×20mL
Pre-Mixed Detection Reagent A 1×120μL Assay Diluent A 1×12mL
Detection Reagent B (PE-SA) 1×120μL Assay Diluent B 1×12mL
Sheath Fluid 1×10mL Wash Buffer (30 × concentrate) 1×20mL
Instruction manual 1

Assay procedure summary

1. Preparation of standards, reagents and samples before the experiment;
2. Add 100μL standard or sample to each well,
    add 10μL magnetic beads, and incubate 90min at 37°C on shaker;
3. Remove liquid on magnetic frame, add 100μL prepared Detection Reagent A. Incubate 60min at 37°C on shaker;
4. Wash plate on magnetic frame for three times;
5. Add 100μL prepared Detection Reagent B, and incubate 30 min at 37°C on shaker;
6. Wash plate on magnetic frame for three times;
7. Add 100μL sheath solution, swirl for 2 minutes, read on the machine.

GIVEAWAYS

INCREMENT SERVICES

Magazine Citations
International Journal of Medical Sciences Glutathione S-Transferase P1 Correlated with Oxidative Stress in Hepatocellular Carcinoma PubMed: PMC3619117
Molecular Medicine Reports Resveratrol ameliorates oxidative stress and inhibits aquaporin 4 expression following rat cerebral ischemia-reperfusion injury PubMed: 26458999
Journal of Chromatography B Set-up and application of an analytical approach for the quality control of purified colostrum as food supplement Pubmed:27341426
Free radical research Circulating cell-free DNA of methylated insulin-like growth factor-binding protein 7 predicts a poor prognosis in hepatitis B virus-associated hepatocellular carcinoma … Pubmed:29463155
Catalog No. Related products for research use of Homo sapiens (Human) Organism species Applications (RESEARCH USE ONLY!)
RPC608Hu01 Recombinant Xanthine Dehydrogenase (XDH) Positive Control; Immunogen; SDS-PAGE; WB.
PAC608Hu01 Polyclonal Antibody to Xanthine Dehydrogenase (XDH) WB; IHC; ICC; IP.
MAC608Hu22 Monoclonal Antibody to Xanthine Dehydrogenase (XDH) WB; IHC; ICC; IP.
SEC608Hu ELISA Kit for Xanthine Dehydrogenase (XDH) Enzyme-linked immunosorbent assay for Antigen Detection.
SCC608Hu CLIA Kit for Xanthine Dehydrogenase (XDH) Chemiluminescent immunoassay for Antigen Detection.
LMC608Hu Multiplex Assay Kit for Xanthine Dehydrogenase (XDH) ,etc. by FLIA (Flow Luminescence Immunoassay) FLIA Kit for Antigen Detection.