Multiplex Assay Kit for Advanced Oxidation Protein Products (AOPP) ,etc. by FLIA (Flow Luminescence Immunoassay)

(Note: Up to 8-plex in one testing reaction)

Specificity

This assay has high sensitivity and excellent specificity for detection of Advanced Oxidation Protein Products (AOPP) ,etc. by FLIA (Flow Luminescence Immunoassay).
No significant cross-reactivity or interference between Advanced Oxidation Protein Products (AOPP) ,etc. by FLIA (Flow Luminescence Immunoassay) and analogues was observed.

Recovery

Matrices listed below were spiked with certain level of recombinant Advanced Oxidation Protein Products (AOPP) ,etc. by FLIA (Flow Luminescence Immunoassay) and the recovery rates were calculated by comparing the measured value to the expected amount of Advanced Oxidation Protein Products (AOPP) ,etc. by FLIA (Flow Luminescence Immunoassay) in samples.

Matrix Recovery range (%) Average(%)
serum(n=5) 80-97 81
EDTA plasma(n=5) 95-102 99
heparin plasma(n=5) 80-97 81

Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Advanced Oxidation Protein Products (AOPP) ,etc. by FLIA (Flow Luminescence Immunoassay) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Advanced Oxidation Protein Products (AOPP) ,etc. by FLIA (Flow Luminescence Immunoassay) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Advanced Oxidation Protein Products (AOPP) ,etc. by FLIA (Flow Luminescence Immunoassay) and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Sample 1:2 1:4 1:8 1:16
serum(n=5) 78-99% 78-101% 87-104% 80-98%
EDTA plasma(n=5) 93-102% 80-95% 83-98% 87-96%
heparin plasma(n=5) 99-105% 96-103% 83-97% 78-94%

Stability

The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

Reagents and materials provided

Reagents Quantity Reagents Quantity
96-well plate 1 Plate sealer for 96 wells 4
Pre-Mixed Standard 2 Standard Diluent 1×20mL
Pre-Mixed Magnetic beads (22#:AOPP) 1 Analysis buffer 1×20mL
Pre-Mixed Detection Reagent A 1×120μL Assay Diluent A 1×12mL
Detection Reagent B (PE-SA) 1×120μL Assay Diluent B 1×12mL
Sheath Fluid 1×10mL Wash Buffer (30 × concentrate) 1×20mL
Instruction manual 1

Assay procedure summary

1. Preparation of standards, reagents and samples before the experiment;
2. Add 50μL standard or sample to each well,
    add 10μL magnetic beads,and 50μL Detection Reagent A,incubate 60min at 37°C on shaker;
3. Wash plate on magnetic frame for three times;
4. Add 100μL prepared Detection Reagent B, and incubate 30 min at 37°C on shaker;
5. Wash plate on magnetic frame for three times;
6. Add 100μL sheath solution, swirl for 2 minutes, read on the machine.

GIVEAWAYS

INCREMENT SERVICES

Magazine Citations
Journal of Pain Research Oxidative Stress Contributes to Hyperalgesia in Osteoporotic Mice Pubmed: 32021402
Antioxidants (Basel) Changes in Oxidative Stress, Inflammation, and Muscle Damage Markers Following Diet and Beetroot Juice Supplementation in Elite Fencers Pubmed: 32630279
Oxidative Medicine and Cellular Longevity Allicin Attenuated Advanced Oxidation Protein Product-Induced Oxidative Stress and Mitochondrial Apoptosis in Human Nucleus Pulposus Cells 33381267
Catalog No. Related products for research use of Homo sapiens (Human) Organism species Applications (RESEARCH USE ONLY!)
NPB223Hu01 Native Advanced Oxidation Protein Products (AOPP) Positive Control; Immunogen; SDS-PAGE; WB.
PAB223Hu05 Polyclonal Antibody to Advanced Oxidation Protein Products (AOPP) WB; IHC; ICC; IP.
CEB223Hu ELISA Kit for Advanced Oxidation Protein Products (AOPP) Enzyme-linked immunosorbent assay for Antigen Detection.
LMB223Hu Multiplex Assay Kit for Advanced Oxidation Protein Products (AOPP) ,etc. by FLIA (Flow Luminescence Immunoassay) FLIA Kit for Antigen Detection.
KSB223Hu11 ELISA Kit DIY Materials for Advanced Oxidation Protein Products (AOPP) Main materials for "Do It (ELISA Kit) Yourself".