Multiplex Assay Kit for Estriol (E3) ,etc. by FLIA (Flow Luminescence Immunoassay)

oestriol

(Note: Up to 8-plex in one testing reaction)

Specificity

This assay has high sensitivity and excellent specificity for detection of Estriol (E3) ,etc. by FLIA (Flow Luminescence Immunoassay).
No significant cross-reactivity or interference between Estriol (E3) ,etc. by FLIA (Flow Luminescence Immunoassay) and analogues was observed.

Recovery

Matrices listed below were spiked with certain level of Estriol (E3) ,etc. by FLIA (Flow Luminescence Immunoassay) and the recovery rates were calculated by comparing the measured value to the expected amount of Estriol (E3) ,etc. by FLIA (Flow Luminescence Immunoassay) in samples.

Matrix Recovery range (%) Average(%)
serum(n=5) 79-90 85
EDTA plasma(n=5) 99-105 102
heparin plasma(n=5) 95-102 98
sodium citrate plasma(n=5) 87-98 92

Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Estriol (E3) ,etc. by FLIA (Flow Luminescence Immunoassay) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Estriol (E3) ,etc. by FLIA (Flow Luminescence Immunoassay) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Estriol (E3) ,etc. by FLIA (Flow Luminescence Immunoassay) and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Sample 1:2 1:4 1:8 1:16
serum(n=5) 96-105% 89-103% 86-96% 92-99%
EDTA plasma(n=5) 80-90% 80-98% 91-99% 91-104%
heparin plasma(n=5) 93-102% 93-105% 80-91% 78-89%
sodium citrate plasma(n=5) 80-103% 87-99% 89-103% 82-102%

Stability

The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

Reagents and materials provided

Reagents Quantity Reagents Quantity
96-well plate 1 Plate sealer for 96 wells 4
Pre-Mixed Standard 2 Standard Diluent 1×20mL
Pre-Mixed Magnetic beads (22#:E3) 1 Analysis buffer 1×20mL
Pre-Mixed Detection Reagent A 1×120μL Assay Diluent A 1×12mL
Detection Reagent B (PE-SA) 1×120μL Assay Diluent B 1×12mL
Sheath Fluid 1×10mL Wash Buffer (30 × concentrate) 1×20mL
Instruction manual 1

Assay procedure summary

1. Preparation of standards, reagents and samples before the experiment;
2. Add 50μL standard or sample to each well,
    add 10μL magnetic beads,and 50μL Detection Reagent A,incubate 60min at 37°C on shaker;
3. Wash plate on magnetic frame for three times;
4. Add 100μL prepared Detection Reagent B, and incubate 30 min at 37°C on shaker;
5. Wash plate on magnetic frame for three times;
6. Add 100μL sheath solution, swirl for 2 minutes, read on the machine.

GIVEAWAYS

INCREMENT SERVICES

Magazine Citations
European journal of pharmacology Osteoprotective effects of salidroside in ovariectomized mice and diabetic mice pubmed:29242120
Catalog No. Related products for research use of Pan-species (General) Organism species Applications (RESEARCH USE ONLY!)
CPA455Ge11 BSA Conjugated Estriol (E3) Immunogen; SDS-PAGE; WB.
CPA455Ge21 OVA Conjugated Estriol (E3) Immunogen; SDS-PAGE; WB.
CPA455Ge31 KLH Conjugated Estriol (E3) Immunogen; SDS-PAGE; WB.
PAA455Ge01 Polyclonal Antibody to Estriol (E3) ELISA, CLIA. / IHC-Fr, ICC, IP (predicted).
MAA455Ge22 Monoclonal Antibody to Estriol (E3) ELISA, CLIA. / IHC-Fr, ICC, IP (predicted).
MAA455Ge21 Monoclonal Antibody to Estriol (E3) ELISA, CLIA. / IHC-Fr, ICC, IP (predicted).
CEA455Ge ELISA Kit for Estriol (E3) Enzyme-linked immunosorbent assay for Antigen Detection.
IEA455Ge Instant ELISA Kit for Estriol (E3) Enzyme-linked immunosorbent assay for Antigen Detection.
CCA455Ge CLIA Kit for Estriol (E3) Chemiluminescent immunoassay for Antigen Detection.
LMA455Ge Multiplex Assay Kit for Estriol (E3) ,etc. by FLIA (Flow Luminescence Immunoassay) FLIA Kit for Antigen Detection.
KSA455Ge11 ELISA Kit DIY Materials for Estriol (E3) Main materials for "Do It (ELISA Kit) Yourself".