Multiplex Assay Kit for Myelin Associated Glycoprotein (MAG) ,etc. by FLIA (Flow Luminescence Immunoassay)

GMA; S-MAG; Siglec-4; SIGLEC4A; Sialic Acid Binding Ig-Like Lectin 4A

(Note: Up to 8-plex in one testing reaction)

Specificity

This assay has high sensitivity and excellent specificity for detection of Myelin Associated Glycoprotein (MAG).
No significant cross-reactivity or interference between Myelin Associated Glycoprotein (MAG) and analogues was observed.

Recovery

Matrices listed below were spiked with certain level of recombinant Myelin Associated Glycoprotein (MAG) and the recovery rates were calculated by comparing the measured value to the expected amount of Myelin Associated Glycoprotein (MAG) in samples.

Matrix Recovery range (%) Average(%)
serum(n=5) 81-102 90
EDTA plasma(n=5) 84-94 90
heparin plasma(n=5) 97-105 101

Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Myelin Associated Glycoprotein (MAG) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Myelin Associated Glycoprotein (MAG) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Myelin Associated Glycoprotein (MAG) and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Sample 1:2 1:4 1:8 1:16
serum(n=5) 93-101% 87-95% 80-96% 98-105%
EDTA plasma(n=5) 81-94% 78-96% 83-91% 92-105%
heparin plasma(n=5) 79-91% 83-96% 93-101% 85-98%

Stability

The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

Reagents and materials provided

Reagents Quantity Reagents Quantity
96-well plate 1 Plate sealer for 96 wells 4
Pre-Mixed Standard 2 Standard Diluent 1×20mL
Pre-Mixed Magnetic beads (22#:MAG) 1 Analysis buffer 1×20mL
Pre-Mixed Detection Reagent A 1×120μL Assay Diluent A 1×12mL
Detection Reagent B (PE-SA) 1×120μL Assay Diluent B 1×12mL
Sheath Fluid 1×10mL Wash Buffer (30 × concentrate) 1×20mL
Instruction manual 1

Assay procedure summary

1. Add 200μL analysis buffer solution to each well of the plate for pre-wetting;
2. Preparation of standards, reagents and samples before the experiment;
3. Add 100μL standard or sample to each well, add 10μL magnetic beads, and incubate 120 minutes protect from light at 25°C on shaker;
4. Remove liquid on magnetic frame, add 100μL prepared Detection Reagent A. Incubate 60 minutes protect from light at 25°C on shaker;
5. Remove liquid on magnetic frame, add 100μL prepared Detection Reagent B, and incubate 30 minutes protect from light at 25°C on shaker;
6. Wash plate once on magnetic frame;
7. Add 100μL sheath fluid, swirl for 10 minutes, read on the machine.

GIVEAWAYS

INCREMENT SERVICES

Magazine Citations
Journal of Materials Science Effect of bimodal pore structure on the bioactivity of poly(lactic-co-glycolic acid)/poly(γ-glutamic acid)/Pluronic 17R4 nerve conduits s10853-016-0729-3
Journal of Materials Science Effect of bimodal pore structure on the bioactivity of poly (lactic-co-glycolic acid)/poly (γ-glutamic acid)/Pluronic 17R4 nerve conduits NRF-2015K2A2A4000128
Catalog No. Related products for research use of Homo sapiens (Human) Organism species Applications (RESEARCH USE ONLY!)
RPA422Hu02 Recombinant Myelin Associated Glycoprotein (MAG) Positive Control; Immunogen; SDS-PAGE; WB.
RPA422Hu01 Recombinant Myelin Associated Glycoprotein (MAG) Positive Control; Immunogen; SDS-PAGE; WB.
PAA422Hu02 Polyclonal Antibody to Myelin Associated Glycoprotein (MAG) WB
PAA422Hu01 Polyclonal Antibody to Myelin Associated Glycoprotein (MAG) WB; IHC
MAA422Hu22 Monoclonal Antibody to Myelin Associated Glycoprotein (MAG) WB; IHC; ICC; IP.
MAA422Hu21 Monoclonal Antibody to Myelin Associated Glycoprotein (MAG) WB
SEA422Hu ELISA Kit for Myelin Associated Glycoprotein (MAG) Enzyme-linked immunosorbent assay for Antigen Detection.
AEA422Hu ELISA Kit for Anti-Myelin Associated Glycoprotein Antibody (Anti-MAG) Enzyme-linked immunosorbent assay for Antibody Detection.
LMA422Hu Multiplex Assay Kit for Myelin Associated Glycoprotein (MAG) ,etc. by FLIA (Flow Luminescence Immunoassay) FLIA Kit for Antigen Detection.
CMA422Hu Multiplex Assay Kit for Myelin Associated Glycoprotein (MAG) ,etc. by CBA (Cytometric Bead Array) FLIA Kit for Antigen Detection.
KSA422Hu01 ELISA Kit DIY Materials for Myelin Associated Glycoprotein (MAG) Main materials for "Do It (ELISA Kit) Yourself".