Multiplex Assay Kit for Interleukin 18 Binding Protein (IL18BP) ,etc. by CBA (Cytometric Bead Array)

IL18BPa; MC51L-53L-54L Homolog Gene Product; Tadekinig-alfa

Specificity

This assay has high sensitivity and excellent specificity for detection of Interleukin 18 Binding Protein (IL18BP) ,etc. by CBA (Cytometric Bead Array).
No significant cross-reactivity or interference between Interleukin 18 Binding Protein (IL18BP) ,etc. by CBA (Cytometric Bead Array) and analogues was observed.

Recovery

Matrices listed below were spiked with certain level of recombinant Interleukin 18 Binding Protein (IL18BP) ,etc. by CBA (Cytometric Bead Array) and the recovery rates were calculated by comparing the measured value to the expected amount of Interleukin 18 Binding Protein (IL18BP) ,etc. by CBA (Cytometric Bead Array) in samples.

Matrix Recovery range (%) Average(%)
serum(n=5) 87-94 90
EDTA plasma(n=5) 78-95 90
heparin plasma(n=5) 95-105 101

Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Interleukin 18 Binding Protein (IL18BP) ,etc. by CBA (Cytometric Bead Array) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Interleukin 18 Binding Protein (IL18BP) ,etc. by CBA (Cytometric Bead Array) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Interleukin 18 Binding Protein (IL18BP) ,etc. by CBA (Cytometric Bead Array) and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Sample 1:2 1:4 1:8 1:16
serum(n=5) 91-105% 78-89% 83-97% 88-101%
EDTA plasma(n=5) 80-101% 94-103% 88-95% 85-93%
heparin plasma(n=5) 78-89% 79-94% 93-102% 93-101%

Stability

The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

Reagents and materials provided

Reagents Quantity Reagents Quantity

Assay procedure summary

1. Add 200μL analysis buffer solution to each well of the plate for pre-wetting;
2. Preparation of standards, reagents and samples before the experiment;
3. Add 100μL standard or sample to each well, add 10μL magnetic beads, and incubate 120 minutes protect from light at 25°C on shaker;
4. Remove liquid on magnetic frame, add 100μL prepared Detection Reagent A. Incubate 60 minutes protect from light at 25°C on shaker;
5. Remove liquid on magnetic frame, add 100μL prepared Detection Reagent B, and incubate 30 minutes protect from light at 25°C on shaker;
6. Wash plate once on magnetic frame;
7. Add 100μL sheath fluid, swirl for 10 minutes, read on the machine.

GIVEAWAYS

INCREMENT SERVICES

Magazine Citations
Acta Physiologica IL-18 neutralization during alveolar hypoxia improves left ventricular diastolic function in mice Pubmed:25182570
Journal of Molecular Cell Biology Role of IL‐18 in atopic asthma is determined by balance of IL‐18/IL‐18BP/IL‐18R Pubmed:28922563
Scandinavian Journal of Immunology  Altered expression of IL‐18 binding protein and IL‐18 receptor in basophils and mast cells of asthma patients Pubmed:29505668
Scientific Reports IL-18 binding protein (IL-18BP) as a novel radiation countermeasure after radiation exposure in mice Pubmed: 33129789
Catalog No. Related products for research use of Rattus norvegicus (Rat) Organism species Applications (RESEARCH USE ONLY!)
RPH821Ra01 Recombinant Interleukin 18 Binding Protein (IL18BP) Positive Control; Immunogen; SDS-PAGE; WB.
PAH821Ra01 Polyclonal Antibody to Interleukin 18 Binding Protein (IL18BP) IHC
MAH821Ra21 Monoclonal Antibody to Interleukin 18 Binding Protein (IL18BP) WB; IHC; ICC; IP.
SEH821Ra ELISA Kit for Interleukin 18 Binding Protein (IL18BP) Enzyme-linked immunosorbent assay for Antigen Detection.
SCH821Ra CLIA Kit for Interleukin 18 Binding Protein (IL18BP) Chemiluminescent immunoassay for Antigen Detection.
LMH821Ra Multiplex Assay Kit for Interleukin 18 Binding Protein (IL18BP) ,etc. by FLIA (Flow Luminescence Immunoassay) FLIA Kit for Antigen Detection.
CMH821Ra Multiplex Assay Kit for Interleukin 18 Binding Protein (IL18BP) ,etc. by CBA (Cytometric Bead Array) FLIA Kit for Antigen Detection.