Multiplex Assay Kit for Uromodulin (UMOD) ,etc. by CBA (Cytometric Bead Array)

THP; Uromucoid,Tamm-Horsfall Glycoprotein; Tamm-Horsfall urinary glycoprotein

Specificity

This assay has high sensitivity and excellent specificity for detection of Uromodulin (UMOD) ,etc. by CBA (Cytometric Bead Array).
No significant cross-reactivity or interference between Uromodulin (UMOD) ,etc. by CBA (Cytometric Bead Array) and analogues was observed.

Recovery

Matrices listed below were spiked with certain level of recombinant Uromodulin (UMOD) ,etc. by CBA (Cytometric Bead Array) and the recovery rates were calculated by comparing the measured value to the expected amount of Uromodulin (UMOD) ,etc. by CBA (Cytometric Bead Array) in samples.

Matrix Recovery range (%) Average(%)
serum(n=5) 96-105 101
EDTA plasma(n=5) 95-102 98
heparin plasma(n=5) 84-93 89

Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Uromodulin (UMOD) ,etc. by CBA (Cytometric Bead Array) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Uromodulin (UMOD) ,etc. by CBA (Cytometric Bead Array) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Uromodulin (UMOD) ,etc. by CBA (Cytometric Bead Array) and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Sample 1:2 1:4 1:8 1:16
serum(n=5) 79-96% 88-97% 80-92% 85-99%
EDTA plasma(n=5) 84-101% 94-105% 90-103% 86-101%
heparin plasma(n=5) 96-105% 80-101% 90-97% 81-95%

Stability

The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

Reagents and materials provided

Reagents Quantity Reagents Quantity

Assay procedure summary

1. Add 200μL analysis buffer solution to each well of the plate for pre-wetting;
2. Preparation of standards, reagents and samples before the experiment;
3. Add 100μL standard or sample to each well, add 10μL magnetic beads, and incubate 120 minutes protect from light at 25°C on shaker;
4. Remove liquid on magnetic frame, add 100μL prepared Detection Reagent A. Incubate 60 minutes protect from light at 25°C on shaker;
5. Remove liquid on magnetic frame, add 100μL prepared Detection Reagent B, and incubate 30 minutes protect from light at 25°C on shaker;
6. Wash plate once on magnetic frame;
7. Add 100μL sheath fluid, swirl for 10 minutes, read on the machine.

GIVEAWAYS

INCREMENT SERVICES

Magazine Citations
Evid Based Complement Alternat Med Quercetin Protects against Cadmium-Induced Renal Uric Acid Transport System Alteration and Lipid Metabolism Disorder in Rats PubMed: PMC3368504
Journal of Korean Medical Science Fate of Neutrophils during the Recovery Phase of Ischemia/Reperfusion Induced Acute Kidney Injury pubmed:28875605
Eberhard Karls University Experimental nephrotic syndrome in mice with an inducible deletion of podocin and the role of plasminogen in sodium retention
Catalog No. Related products for research use of Homo sapiens (Human) Organism species Applications (RESEARCH USE ONLY!)
EPG918Hu61 Eukaryotic Uromodulin (UMOD) Positive Control; Immunogen; SDS-PAGE; WB.
RPG918Hu01 Recombinant Uromodulin (UMOD) Positive Control; Immunogen; SDS-PAGE; WB.
CPG918Hu21 OVA Conjugated Uromodulin (UMOD) Immunogen; SDS-PAGE; WB.
PAG918Hu08 Polyclonal Antibody to Uromodulin (UMOD) WB
PAG918Hu01 Polyclonal Antibody to Uromodulin (UMOD) WB; ICC/IF
LAG918Hu71 Biotin-Linked Polyclonal Antibody to Uromodulin (UMOD) WB; IHC; ICC.
HEG918Hu High Sensitive ELISA Kit for Uromodulin (UMOD) Enzyme-linked immunosorbent assay for Antigen Detection.
SEG918Hu ELISA Kit for Uromodulin (UMOD) Enzyme-linked immunosorbent assay for Antigen Detection.
LMG918Hu Multiplex Assay Kit for Uromodulin (UMOD) ,etc. by FLIA (Flow Luminescence Immunoassay) FLIA Kit for Antigen Detection.
CMG918Hu Multiplex Assay Kit for Uromodulin (UMOD) ,etc. by CBA (Cytometric Bead Array) FLIA Kit for Antigen Detection.