Multiplex Assay Kit for Histamine Receptor H1 (HRH1) ,etc. by CBA (Cytometric Bead Array)

H1-R; hisH1

Specificity

This assay has high sensitivity and excellent specificity for detection of Histamine Receptor H1 (HRH1) ,etc. by CBA (Cytometric Bead Array).
No significant cross-reactivity or interference between Histamine Receptor H1 (HRH1) ,etc. by CBA (Cytometric Bead Array) and analogues was observed.

Recovery

Matrices listed below were spiked with certain level of recombinant Histamine Receptor H1 (HRH1) ,etc. by CBA (Cytometric Bead Array) and the recovery rates were calculated by comparing the measured value to the expected amount of Histamine Receptor H1 (HRH1) ,etc. by CBA (Cytometric Bead Array) in samples.

Matrix Recovery range (%) Average(%)
serum(n=5) 98-105 101
EDTA plasma(n=5) 85-99 90
heparin plasma(n=5) 78-98 93

Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Histamine Receptor H1 (HRH1) ,etc. by CBA (Cytometric Bead Array) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Histamine Receptor H1 (HRH1) ,etc. by CBA (Cytometric Bead Array) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Histamine Receptor H1 (HRH1) ,etc. by CBA (Cytometric Bead Array) and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Sample 1:2 1:4 1:8 1:16
serum(n=5) 89-97% 78-99% 92-105% 85-92%
EDTA plasma(n=5) 92-102% 90-103% 86-96% 79-88%
heparin plasma(n=5) 96-103% 82-93% 93-102% 84-95%

Stability

The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

Reagents and materials provided

Reagents Quantity Reagents Quantity

Assay procedure summary

1. Add 200μL analysis buffer solution to each well of the plate for pre-wetting;
2. Preparation of standards, reagents and samples before the experiment;
3. Add 100μL standard or sample to each well, add 10μL magnetic beads, and incubate 120 minutes protect from light at 25°C on shaker;
4. Remove liquid on magnetic frame, add 100μL prepared Detection Reagent A. Incubate 60 minutes protect from light at 25°C on shaker;
5. Remove liquid on magnetic frame, add 100μL prepared Detection Reagent B, and incubate 30 minutes protect from light at 25°C on shaker;
6. Wash plate once on magnetic frame;
7. Add 100μL sheath fluid, swirl for 10 minutes, read on the machine.

GIVEAWAYS

INCREMENT SERVICES

Magazine Citations
annal of medicine and surgery Laparoscopic procedures impact on mast cell mediators, extracellular matrix and adhesion scoring system in rats Pubmed: 32963775
Catalog No. Related products for research use of Rattus norvegicus (Rat) Organism species Applications (RESEARCH USE ONLY!)
RPG037Ra01 Recombinant Histamine Receptor H1 (HRH1) Positive Control; Immunogen; SDS-PAGE; WB.
PAG037Ra01 Polyclonal Antibody to Histamine Receptor H1 (HRH1) WB; IHC; ICC; IP.
MAG037Ra21 Monoclonal Antibody to Histamine Receptor H1 (HRH1) WB; IHC; ICC; IP.
SEG037Ra ELISA Kit for Histamine Receptor H1 (HRH1) Enzyme-linked immunosorbent assay for Antigen Detection.
LMG037Ra Multiplex Assay Kit for Histamine Receptor H1 (HRH1) ,etc. by FLIA (Flow Luminescence Immunoassay) FLIA Kit for Antigen Detection.
CMG037Ra Multiplex Assay Kit for Histamine Receptor H1 (HRH1) ,etc. by CBA (Cytometric Bead Array) FLIA Kit for Antigen Detection.