Multiplex Assay Kit for Chorionic Gonadotropin (CG) ,etc. by CBA (Cytometric Bead Array)

hCG; Human Chorionic Gonadotrophin

Specificity

This assay has high sensitivity and excellent specificity for detection of Chorionic Gonadotropin (CG) ,etc. by CBA (Cytometric Bead Array).
No significant cross-reactivity or interference between Chorionic Gonadotropin (CG) ,etc. by CBA (Cytometric Bead Array) and analogues was observed.

Recovery

Matrices listed below were spiked with certain level of recombinant Chorionic Gonadotropin (CG) ,etc. by CBA (Cytometric Bead Array) and the recovery rates were calculated by comparing the measured value to the expected amount of Chorionic Gonadotropin (CG) ,etc. by CBA (Cytometric Bead Array) in samples.

Matrix Recovery range (%) Average(%)
serum(n=5) 94-105 102
EDTA plasma(n=5) 98-105 101
heparin plasma(n=5) 78-102 81

Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Chorionic Gonadotropin (CG) ,etc. by CBA (Cytometric Bead Array) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Chorionic Gonadotropin (CG) ,etc. by CBA (Cytometric Bead Array) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Chorionic Gonadotropin (CG) ,etc. by CBA (Cytometric Bead Array) and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Sample 1:2 1:4 1:8 1:16
serum(n=5) 94-101% 88-101% 80-101% 98-105%
EDTA plasma(n=5) 94-102% 93-101% 98-105% 79-105%
heparin plasma(n=5) 98-105% 83-99% 80-99% 86-97%

Stability

The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

Reagents and materials provided

Reagents Quantity Reagents Quantity

Assay procedure summary

1. Add 200μL analysis buffer solution to each well of the plate for pre-wetting;
2. Preparation of standards, reagents and samples before the experiment;
3. Add 100μL standard or sample to each well, add 10μL magnetic beads, and incubate 120 minutes protect from light at 25°C on shaker;
4. Remove liquid on magnetic frame, add 100μL prepared Detection Reagent A. Incubate 60 minutes protect from light at 25°C on shaker;
5. Remove liquid on magnetic frame, add 100μL prepared Detection Reagent B, and incubate 30 minutes protect from light at 25°C on shaker;
6. Wash plate once on magnetic frame;
7. Add 100μL sheath fluid, swirl for 10 minutes, read on the machine.

GIVEAWAYS

INCREMENT SERVICES

Magazine Citations
Placenta The role of Sep (O-phosphoserine) tRNA: Sec (selenocysteine) synthase (SEPSECS) in proliferation, apoptosis and hormone secretion of trophoblast cells Pubmed: 23966103
International Journal of Nanomedicine Uptake and transport of pullulan acetate nanoparticles in the BeWo b30 placental barrier cell model Pubmed:30034233
Thyroid Defending effects of iodide transfer in placental barrier against maternal iodine deficiency Pubmed: 32791891
Catalog No. Related products for research use of Homo sapiens (Human) Organism species Applications (RESEARCH USE ONLY!)
PAD216Hu01 Polyclonal Antibody to Chorionic Gonadotropin (CG) WB; IHC; ICC; IP.
LAD216Hu71 Biotin-Linked Polyclonal Antibody to Chorionic Gonadotropin (CG) WB; IHC; ICC.
SED216Hu ELISA Kit for Chorionic Gonadotropin (CG) Enzyme-linked immunosorbent assay for Antigen Detection.
SCD216Hu CLIA Kit for Chorionic Gonadotropin (CG) Chemiluminescent immunoassay for Antigen Detection.
LMD216Hu Multiplex Assay Kit for Chorionic Gonadotropin (CG) ,etc. by FLIA (Flow Luminescence Immunoassay) FLIA Kit for Antigen Detection.
CMD216Hu Multiplex Assay Kit for Chorionic Gonadotropin (CG) ,etc. by CBA (Cytometric Bead Array) FLIA Kit for Antigen Detection.