Multiplex Assay Kit for Ferritin, Light Polypeptide (FTL) ,etc. by CBA (Cytometric Bead Array)

Ferritin Light Chain; Ferritin Light Polypeptide-Like 3; L Apoferritin; Ferritin L Subunit

Specificity

This assay has high sensitivity and excellent specificity for detection of Ferritin, Light Polypeptide (FTL) ,etc. by CBA (Cytometric Bead Array).
No significant cross-reactivity or interference between Ferritin, Light Polypeptide (FTL) ,etc. by CBA (Cytometric Bead Array) and analogues was observed.

Recovery

Matrices listed below were spiked with certain level of recombinant Ferritin, Light Polypeptide (FTL) ,etc. by CBA (Cytometric Bead Array) and the recovery rates were calculated by comparing the measured value to the expected amount of Ferritin, Light Polypeptide (FTL) ,etc. by CBA (Cytometric Bead Array) in samples.

Matrix Recovery range (%) Average(%)
serum(n=5) 99-105 102
EDTA plasma(n=5) 96-103 99
heparin plasma(n=5) 79-97 88

Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Ferritin, Light Polypeptide (FTL) ,etc. by CBA (Cytometric Bead Array) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Ferritin, Light Polypeptide (FTL) ,etc. by CBA (Cytometric Bead Array) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Ferritin, Light Polypeptide (FTL) ,etc. by CBA (Cytometric Bead Array) and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Sample 1:2 1:4 1:8 1:16
serum(n=5) 97-104% 79-102% 84-93% 82-103%
EDTA plasma(n=5) 79-98% 92-105% 90-98% 96-105%
heparin plasma(n=5) 85-98% 95-103% 95-103% 90-98%

Stability

The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

Reagents and materials provided

Reagents Quantity Reagents Quantity

Assay procedure summary

1. Add 200μL analysis buffer solution to each well of the plate for pre-wetting;
2. Preparation of standards, reagents and samples before the experiment;
3. Add 100μL standard or sample to each well, add 10μL magnetic beads, and incubate 120 minutes protect from light at 25°C on shaker;
4. Remove liquid on magnetic frame, add 100μL prepared Detection Reagent A. Incubate 60 minutes protect from light at 25°C on shaker;
5. Remove liquid on magnetic frame, add 100μL prepared Detection Reagent B, and incubate 30 minutes protect from light at 25°C on shaker;
6. Wash plate once on magnetic frame;
7. Add 100μL sheath fluid, swirl for 10 minutes, read on the machine.

GIVEAWAYS

INCREMENT SERVICES

Magazine Citations
PLoS One Presence in the Pre-Surgical Fine-Needle Aspiration of Potential Thyroid Biomarkers Previously Identified in the Post-Surgical One PubMed: PMC3765839
Am J Physiol Lung Cell Mol Physiol. Methemoglobin-induced signaling and chemokine responses in human alveolar epithelial cells Pubmed:24142518
Catalog No. Related products for research use of Mus musculus (Mouse) Organism species Applications (RESEARCH USE ONLY!)
RPD020Mu01 Recombinant Ferritin, Light Polypeptide (FTL) Positive Control; Immunogen; SDS-PAGE; WB.
RPD020Mu02 Recombinant Ferritin, Light Polypeptide (FTL) Positive Control; Immunogen; SDS-PAGE; WB.
PAD020Mu02 Polyclonal Antibody to Ferritin, Light Polypeptide (FTL) WB; IHC; ICC; IP.
PAD020Mu01 Polyclonal Antibody to Ferritin, Light Polypeptide (FTL) WB; IHC
SED020Mu ELISA Kit for Ferritin, Light Polypeptide (FTL) Enzyme-linked immunosorbent assay for Antigen Detection.
CMD020Mu Multiplex Assay Kit for Ferritin, Light Polypeptide (FTL) ,etc. by CBA (Cytometric Bead Array) FLIA Kit for Antigen Detection.
LMD020Mu Multiplex Assay Kit for Ferritin, Light Polypeptide (FTL) ,etc. by FLIA (Flow Luminescence Immunoassay) FLIA Kit for Antigen Detection.