Multiplex Assay Kit for Latent Transforming Growth Factor Beta Binding Protein 2 (LTBP2) ,etc. by CBA (Cytometric Bead Array)

C14orf141; LTBP3

Specificity

This assay has high sensitivity and excellent specificity for detection of Latent Transforming Growth Factor Beta Binding Protein 2 (LTBP2) ,etc. by CBA (Cytometric Bead Array).
No significant cross-reactivity or interference between Latent Transforming Growth Factor Beta Binding Protein 2 (LTBP2) ,etc. by CBA (Cytometric Bead Array) and analogues was observed.

Recovery

Matrices listed below were spiked with certain level of recombinant Latent Transforming Growth Factor Beta Binding Protein 2 (LTBP2) ,etc. by CBA (Cytometric Bead Array) and the recovery rates were calculated by comparing the measured value to the expected amount of Latent Transforming Growth Factor Beta Binding Protein 2 (LTBP2) ,etc. by CBA (Cytometric Bead Array) in samples.

Matrix Recovery range (%) Average(%)
serum(n=5) 79-94 82
EDTA plasma(n=5) 97-105 101
heparin plasma(n=5) 91-99 95

Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Latent Transforming Growth Factor Beta Binding Protein 2 (LTBP2) ,etc. by CBA (Cytometric Bead Array) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Latent Transforming Growth Factor Beta Binding Protein 2 (LTBP2) ,etc. by CBA (Cytometric Bead Array) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Latent Transforming Growth Factor Beta Binding Protein 2 (LTBP2) ,etc. by CBA (Cytometric Bead Array) and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Sample 1:2 1:4 1:8 1:16
serum(n=5) 78-101% 90-99% 98-105% 95-104%
EDTA plasma(n=5) 78-92% 94-101% 85-99% 97-105%
heparin plasma(n=5) 89-103% 80-89% 79-99% 91-98%

Stability

The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

Reagents and materials provided

Reagents Quantity Reagents Quantity

Assay procedure summary

1. Add 200μL analysis buffer solution to each well of the plate for pre-wetting;
2. Preparation of standards, reagents and samples before the experiment;
3. Add 100μL standard or sample to each well, add 10μL magnetic beads, and incubate 120 minutes protect from light at 25°C on shaker;
4. Remove liquid on magnetic frame, add 100μL prepared Detection Reagent A. Incubate 60 minutes protect from light at 25°C on shaker;
5. Remove liquid on magnetic frame, add 100μL prepared Detection Reagent B, and incubate 30 minutes protect from light at 25°C on shaker;
6. Wash plate once on magnetic frame;
7. Add 100μL sheath fluid, swirl for 10 minutes, read on the machine.

GIVEAWAYS

INCREMENT SERVICES

Magazine Citations
International Journal of Cancer Osteopontin and latent-TGF β binding-protein 2 as potential diagnostic markers for HBV-related hepatocellular carcinoma Pubmed:24803312
Int J Cancer Osteopontin and latent‐TGF β binding‐protein 2 as potential diagnostic markers for HBV‐related hepatocellular carcinoma PubMed: 24803312
Biology of Blood and Marrow Transplantation Association of Plasma CD163 Concentration with De Novo–Onset Chronic Graft-versus-Host Disease pubmed:28455006
Biology of Blood and Marrow Transplantation Association of plasma CD163 concentration with de novo-onset chronic graft-versus-host disease 10.1016:j.bbmt.2017.04.019
Clinical Science LTBP2 is secreted from lung myofibroblasts and is a potential biomarker for idiopathic pulmonary fibrosis Pubmed:30006483