Multiplex Assay Kit for Cardiotrophin 1 (CT1) ,etc. by CBA (Cytometric Bead Array)

CTF1

Specificity

This assay has high sensitivity and excellent specificity for detection of Cardiotrophin 1 (CT1) ,etc. by CBA (Cytometric Bead Array).
No significant cross-reactivity or interference between Cardiotrophin 1 (CT1) ,etc. by CBA (Cytometric Bead Array) and analogues was observed.

Recovery

Matrices listed below were spiked with certain level of recombinant Cardiotrophin 1 (CT1) ,etc. by CBA (Cytometric Bead Array) and the recovery rates were calculated by comparing the measured value to the expected amount of Cardiotrophin 1 (CT1) ,etc. by CBA (Cytometric Bead Array) in samples.

Matrix Recovery range (%) Average(%)
serum(n=5) 79-96 86
EDTA plasma(n=5) 88-102 92
heparin plasma(n=5) 80-91 84

Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Cardiotrophin 1 (CT1) ,etc. by CBA (Cytometric Bead Array) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Cardiotrophin 1 (CT1) ,etc. by CBA (Cytometric Bead Array) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Cardiotrophin 1 (CT1) ,etc. by CBA (Cytometric Bead Array) and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Sample 1:2 1:4 1:8 1:16
serum(n=5) 84-97% 97-105% 83-104% 80-95%
EDTA plasma(n=5) 92-101% 80-97% 82-94% 92-102%
heparin plasma(n=5) 79-99% 92-102% 91-101% 79-96%

Stability

The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

Reagents and materials provided

Reagents Quantity Reagents Quantity
96-well plate 1 Plate sealer for 96 wells 4
Pre-Mixed Standard 2 Standard Diluent 1×20mL
Pre-Mixed Magnetic beads (22#:CT1) 1 Analysis buffer 1×20mL
Pre-Mixed Detection Reagent A 1×120μL Assay Diluent A 1×12mL
Detection Reagent B (PE-SA) 1×120μL Assay Diluent B 1×12mL
Sheath Fluid 1×10mL Wash Buffer (30 × concentrate) 1×20mL
Instruction manual 1

Assay procedure summary

1. Add 200μL analysis buffer solution to each well of the plate for pre-wetting;
2. Preparation of standards, reagents and samples before the experiment;
3. Add 100μL standard or sample to each well, add 10μL magnetic beads, and incubate 120 minutes protect from light at 25°C on shaker;
4. Remove liquid on magnetic frame, add 100μL prepared Detection Reagent A. Incubate 60 minutes protect from light at 25°C on shaker;
5. Remove liquid on magnetic frame, add 100μL prepared Detection Reagent B, and incubate 30 minutes protect from light at 25°C on shaker;
6. Wash plate once on magnetic frame;
7. Add 100μL sheath fluid, swirl for 10 minutes, read on the machine.

GIVEAWAYS

INCREMENT SERVICES

Magazine Citations
European Journal of Preventive Cardiology Exercise rescued chronic kidney disease by attenuating cardiac hypertrophy through the cardiotrophin-1 -> LIFR/gp 130 -> JAK/STAT3 pathway. PubMed: 23064267
Journal of Hypertension Is There Any Additional Cardiotrophin 1 Increase With Obesity in Hypertensive Patients?: Pp. 12.466 Lww: Source
Journal of Hypertension Cardiotrophyn 1 Levels in Nondipper Hypertensive Patients: Pp. 12.464 Lww: Source
American Journal of Hypertension Association Between Cardiotrophin 1 Levels and Central Blood Pressure in Untreated Patients With Essential Hypertension Pubmed: 24401751
Scientific Reports Biomarkers and Echocardiographic Predictors of Myocardial Dysfunction in Patients with Hypertension PubMed: 25747153
Int J Mol Sci The Multi-Biomarker Approach for Heart Failure in Patients with Hypertension PubMed: 25984599
Food & Function Pterostilbene improves glycaemic control in rats fed an obesogenic diet: involvement of skeletal muscle and liver PubMed: 25998070
International journal of cardiology Differences in biochemical and genetic biomarkers in patients with heart failure of various etiologies. pubmed:27448535
Soldan sağa şantlı doğumsal kalp hastalıklı çocuklarda şant miktarı ile kardiyotrofin–1 ilişkisi Doi: 10.5505/deutfd.2018.25338.30074
Annals of Nutrition and Metabolism Effect of Salt Intake on the Serum Cardiotrophin-1 Levels in Chinese Adults Pubmed: 30408810
Atherosclerotic diseases Cardiotrophin-1 and leptin as cardiovascular risk markers in male patients with obstructive sleep apnea syndrome Pubmed: 30775602
Catalog No. Related products for research use of Mus musculus (Mouse) Organism species Applications (RESEARCH USE ONLY!)
RPA810Mu01 Recombinant Cardiotrophin 1 (CT1) Positive Control; Immunogen; SDS-PAGE; WB.
APA810Mu01 Active Cardiotrophin 1 (CT1) Cell culture; Activity Assays.
RPA810Mu02 Recombinant Cardiotrophin 1 (CT1) Positive Control; Immunogen; SDS-PAGE; WB.
PAA810Mu01 Polyclonal Antibody to Cardiotrophin 1 (CT1) Testing in progress
PAA810Mu02 Polyclonal Antibody to Cardiotrophin 1 (CT1) WB; IHC; ICC; IP.
SEA810Mu ELISA Kit for Cardiotrophin 1 (CT1) Enzyme-linked immunosorbent assay for Antigen Detection.
CMA810Mu Multiplex Assay Kit for Cardiotrophin 1 (CT1) ,etc. by CBA (Cytometric Bead Array) FLIA Kit for Antigen Detection.
LMA810Mu Multiplex Assay Kit for Cardiotrophin 1 (CT1) ,etc. by FLIA (Flow Luminescence Immunoassay) FLIA Kit for Antigen Detection.