Multiplex Assay Kit for Activin AB (ACVAB) ,etc. by CBA (Cytometric Bead Array)

ACV-AB; Activin Beta A Beta B Homodimer

Specificity

This assay has high sensitivity and excellent specificity for detection of Activin AB (ACVAB) ,etc. by CBA (Cytometric Bead Array).
No significant cross-reactivity or interference between Activin AB (ACVAB) ,etc. by CBA (Cytometric Bead Array) and analogues was observed.

Recovery

Matrices listed below were spiked with certain level of recombinant Activin AB (ACVAB) ,etc. by CBA (Cytometric Bead Array) and the recovery rates were calculated by comparing the measured value to the expected amount of Activin AB (ACVAB) ,etc. by CBA (Cytometric Bead Array) in samples.

Matrix Recovery range (%) Average(%)
serum(n=5) 84-102 99
EDTA plasma(n=5) 80-94 88
heparin plasma(n=5) 97-105 101

Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Activin AB (ACVAB) ,etc. by CBA (Cytometric Bead Array) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Activin AB (ACVAB) ,etc. by CBA (Cytometric Bead Array) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Activin AB (ACVAB) ,etc. by CBA (Cytometric Bead Array) and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Sample 1:2 1:4 1:8 1:16
serum(n=5) 83-99% 86-94% 84-105% 88-96%
EDTA plasma(n=5) 80-104% 88-102% 91-105% 86-93%
heparin plasma(n=5) 83-101% 81-104% 80-89% 92-99%

Stability

The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

Reagents and materials provided

Reagents Quantity Reagents Quantity

Assay procedure summary

1. Add 200μL analysis buffer solution to each well of the plate for pre-wetting;
2. Preparation of standards, reagents and samples before the experiment;
3. Add 100μL standard or sample to each well, add 10μL magnetic beads, and incubate 120 minutes protect from light at 25°C on shaker;
4. Remove liquid on magnetic frame, add 100μL prepared Detection Reagent A. Incubate 60 minutes protect from light at 25°C on shaker;
5. Remove liquid on magnetic frame, add 100μL prepared Detection Reagent B, and incubate 30 minutes protect from light at 25°C on shaker;
6. Wash plate once on magnetic frame;
7. Add 100μL sheath fluid, swirl for 10 minutes, read on the machine.

GIVEAWAYS

INCREMENT SERVICES

Magazine Citations
Endocrinology TGFβ Superfamily Members Mediate Androgen Deprivation Therapy-Induced Obese Frailty in Male Mice pubmed:27611336
BMC Cancer. Activins and their related proteins in colon carcinogenesis: insights from early and advanced azoxymethane rat models of colon cancer. pubmed:27835986
Pathology & Oncology Research Profiling Activins and Follistatin in Colorectal Cancer According to Clinical Stage, Tumour Sidedness and Smad4 Status 34867090
Catalog No. Related products for research use of Homo sapiens (Human) Organism species Applications (RESEARCH USE ONLY!)
RPA158Hu01 Recombinant Activin AB (ACVAB) Positive Control; Immunogen; SDS-PAGE; WB.
PAA158Hu01 Polyclonal Antibody to Activin AB (ACVAB) WB
SEA158Hu ELISA Kit for Activin AB (ACVAB) Enzyme-linked immunosorbent assay for Antigen Detection.
LMA158Hu Multiplex Assay Kit for Activin AB (ACVAB) ,etc. by FLIA (Flow Luminescence Immunoassay) FLIA Kit for Antigen Detection.
CMA158Hu Multiplex Assay Kit for Activin AB (ACVAB) ,etc. by CBA (Cytometric Bead Array) FLIA Kit for Antigen Detection.