CLIA Kit for Polyethylene Glycol (PEG)

PEO; POE; Polyethylene oxide; Polyoxyethylene; Carbowax; GoLYTELY; GlycoLax; Fortrans; TriLyte; Colyte; Halflytely; Macrogol; MiraLAX; MoviPrep

Specificity

This assay has high sensitivity and excellent specificity for detection of Polyethylene Glycol (PEG).
No significant cross-reactivity or interference between Polyethylene Glycol (PEG) and analogues was observed.

Recovery

Matrices listed below were spiked with certain level of recombinant Polyethylene Glycol (PEG) and the recovery rates were calculated by comparing the measured value to the expected amount of Polyethylene Glycol (PEG) in samples.

Matrix Recovery range (%) Average(%)
serum(n=5) 93-103 99
EDTA plasma(n=5) 93-102 98
heparin plasma(n=5) 86-95 92

Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Polyethylene Glycol (PEG) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Polyethylene Glycol (PEG) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Polyethylene Glycol (PEG) and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Sample 1:2 1:4 1:8 1:16
serum(n=5) 90-98% 97-105% 82-90% 78-89%
EDTA plasma(n=5) 94-105% 87-95% 85-93% 80-88%
heparin plasma(n=5) 78-91% 83-91% 90-99% 85-101%

Stability

The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

Reagents and materials provided

Reagents Quantity Reagents Quantity
Pre-coated, ready to use 96-well strip plate 1 Plate sealer for 96 wells 4
Standard 2 Standard Diluent 1×20mL
Detection Reagent A 1×120µL Assay Diluent A 1×12mL
Detection Reagent B 1×120µL Assay Diluent B 1×12mL
Substrate A 1×10mL Substrate B 1×2mL
Wash Buffer (30 × concentrate) 1×20mL Instruction manual 1

Assay procedure summary

1. Prepare all reagents, samples and standards;
2. Add 50µL standard or sample to each well.
    And then add 50µL prepared Detection Reagent A immediately.
    Shake and mix. Incubate 1 hour at 37°C;
3. Aspirate and wash 3 times;
4. Add 100µL prepared Detection Reagent B. Incubate 30 minutes at 37°C;
5. Aspirate and wash 5 times;
6. Add 100µL Substrate Solution. Incubate 10 minutes at 37°C;
7. Read RLU value immediately.

GIVEAWAYS

INCREMENT SERVICES

Catalog No. Related products for research use of Pan-species (General) Organism species Applications (RESEARCH USE ONLY!)
CPX163Ge11 BSA Conjugated Polyethylene Glycol (PEG) Immunogen; SDS-PAGE; WB.
CPX163Ge21 OVA Conjugated Polyethylene Glycol (PEG) Immunogen; SDS-PAGE; WB.
PAX163Ge01 Polyclonal Antibody to Polyethylene Glycol (PEG) ELISA, CLIA. / IHC-Fr, ICC, IP (predicted).
CEX163Ge ELISA Kit for Polyethylene Glycol (PEG) Enzyme-linked immunosorbent assay for Antigen Detection.
CCX163Ge CLIA Kit for Polyethylene Glycol (PEG) Chemiluminescent immunoassay for Antigen Detection.
LMX163Ge Multiplex Assay Kit for Polyethylene Glycol (PEG) ,etc. by FLIA (Flow Luminescence Immunoassay) FLIA Kit for Antigen Detection.
KSX163Ge11 ELISA Kit DIY Materials for Polyethylene Glycol (PEG) Main materials for "Do It(ELISA Kit) Yourself"