ELISA Kit for Anti-Coagulation Factor II Antibody (Anti-F2)
FII; TM; PT; Thrombin; Prothrombin; Pro-Thrombin; Activation peptide fragment 1; Activation peptide fragment 2
- Product No.AEA820Hu
- Organism SpeciesHomo sapiens (Human) Same name, Different species.
- Sample Typeserum, plasma and other biological fluids
- Test MethodIndirect ELISA
- Assay Length2h, 30min
- Detection Range3.12-200ng/mL
- SensitivityThe minimum detectable dose of this kit is typically less than 1.14ng/mL.
- DownloadInstruction Manual
- UOM 48T96T 96T*5 96T*10 96T*100
- FOB
US$ 630
For more details, please contact local distributors! US$ 900 US$ 4050 US$ 7650 US$ 63000
Specificity
This assay has high sensitivity and excellent specificity for detection of Anti-Coagulation Factor II Antibody (Anti-F2).
No significant cross-reactivity or interference between Anti-Coagulation Factor II Antibody (Anti-F2) and analogues was observed.
Recovery
Matrices listed below were spiked with certain level of Anti-Coagulation Factor II Antibody (Anti-F2) and the recovery rates were calculated by comparing the measured value to the expected amount of Anti-Coagulation Factor II Antibody (Anti-F2) in samples.
Matrix | Recovery range (%) | Average(%) |
serum(n=5) | 95-105 | 101 |
EDTA plasma(n=5) | 91-105 | 95 |
heparin plasma(n=5) | 80-101 | 97 |
Precision
Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Anti-Coagulation Factor II Antibody (Anti-F2) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Anti-Coagulation Factor II Antibody (Anti-F2) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%
Linearity
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Anti-Coagulation Factor II Antibody (Anti-F2) and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Sample | 1:2 | 1:4 | 1:8 | 1:16 |
serum(n=5) | 91-98% | 83-103% | 82-102% | 99-105% |
EDTA plasma(n=5) | 98-105% | 81-95% | 86-104% | 97-105% |
heparin plasma(n=5) | 78-95% | 78-92% | 82-99% | 80-97% |
Stability
The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.
Reagents and materials provided
Reagents | Quantity | Reagents | Quantity |
Pre-coated, ready to use 96-well strip plate | 1 | Plate sealer for 96 wells | 4 |
Standard | 2 | Standard Diluent | 1×20mL |
Detection Reagent A | 1×120µL | Assay Diluent A | 1×12mL |
TMB Substrate | 1×9mL | Stop Solution | 1×6mL |
Wash Buffer (30 × concentrate) | 1×20mL | Instruction manual | 1 |
Assay procedure summary
1. Prepare all reagents, samples and standards;
2. Add 100µL standard or sample to each well. Incubate 1 hours at 37°C;
3. Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C;
4. Aspirate and wash 5 times;
5. Add 90µL Substrate Solution. Incubate 10-20 minutes at 37°C;
6. Add 50µL Stop Solution. Read at 450nm immediately.
GIVEAWAYS
INCREMENT SERVICES
Magazine | Citations |
Blood Cells Mol Dis | Upregulation of cytokine signalling in platelets increases risk of thrombophilia in severe COVID-19 patients Pubmed:35180460 |
Catalog No. | Related products for research use of Homo sapiens (Human) Organism species | Applications (RESEARCH USE ONLY!) |
NPA820Hu01 | Native Coagulation Factor II (F2) | Positive Control; Immunogen; SDS-PAGE; WB. |
RPA820Hu02 | Recombinant Coagulation Factor II (F2) | Positive Control; Immunogen; SDS-PAGE; WB. |
RPA820Hu03 | Recombinant Coagulation Factor II (F2) | Positive Control; Immunogen; SDS-PAGE; WB. |
RPA820Hu01 | Recombinant Coagulation Factor II (F2) | Positive Control; Immunogen; SDS-PAGE; WB. |
PAA820Hu02 | Polyclonal Antibody to Coagulation Factor II (F2) | WB; IHC; ICC; IP. |
PAA820Hu01 | Polyclonal Antibody to Coagulation Factor II (F2) | WB; IHC; ICC; IP. |
MAA820Hu27 | Monoclonal Antibody to Coagulation Factor II (F2) | WB; IHC; ICC; IP. |
MAA820Hu29 | Monoclonal Antibody to Coagulation Factor II (F2) | WB; IHC; ICC; IP. |
MAA820Hu23 | Monoclonal Antibody to Coagulation Factor II (F2) | WB; IHC; ICC; IP. |
MAA820Hu24 | Monoclonal Antibody to Coagulation Factor II (F2) | WB; IHC; ICC; IP. |
MAA820Hu25 | Monoclonal Antibody to Coagulation Factor II (F2) | WB; IHC; ICC; IP. |
MAA820Hu22 | Monoclonal Antibody to Coagulation Factor II (F2) | WB; IHC; ICC; IP. |
MAA820Hu21 | Monoclonal Antibody to Coagulation Factor II (F2) | WB; IHC; ICC; IP. |
SEA820Hu | ELISA Kit for Coagulation Factor II (F2) | Enzyme-linked immunosorbent assay for Antigen Detection. |
AEA820Hu | ELISA Kit for Anti-Coagulation Factor II Antibody (Anti-F2) | Enzyme-linked immunosorbent assay for Antibody Detection. |
SCA820Hu | CLIA Kit for Coagulation Factor II (F2) | Chemiluminescent immunoassay for Antigen Detection. |