Multiplex Assay Kit for Interleukin 34 (IL34) ,etc. by FLIA (Flow Luminescence Immunoassay)

C16orf77

(Note: Up to 8-plex in one testing reaction)

Specificity

This assay has high sensitivity and excellent specificity for detection of Interleukin 34 (IL34) ,etc. by FLIA (Flow Luminescence Immunoassay).
No significant cross-reactivity or interference between Interleukin 34 (IL34) ,etc. by FLIA (Flow Luminescence Immunoassay) and analogues was observed.

Recovery

Matrices listed below were spiked with certain level of recombinant Interleukin 34 (IL34) ,etc. by FLIA (Flow Luminescence Immunoassay) and the recovery rates were calculated by comparing the measured value to the expected amount of Interleukin 34 (IL34) ,etc. by FLIA (Flow Luminescence Immunoassay) in samples.

Matrix Recovery range (%) Average(%)
serum(n=5) 80-89 81
EDTA plasma(n=5) 86-103 90
heparin plasma(n=5) 90-105 94

Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Interleukin 34 (IL34) ,etc. by FLIA (Flow Luminescence Immunoassay) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Interleukin 34 (IL34) ,etc. by FLIA (Flow Luminescence Immunoassay) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Interleukin 34 (IL34) ,etc. by FLIA (Flow Luminescence Immunoassay) and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Sample 1:2 1:4 1:8 1:16
serum(n=5) 80-91% 91-99% 81-91% 90-98%
EDTA plasma(n=5) 92-105% 84-96% 94-103% 84-97%
heparin plasma(n=5) 88-99% 97-104% 78-91% 89-103%

Stability

The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

Reagents and materials provided

Reagents Quantity Reagents Quantity
96-well plate 1 Plate sealer for 96 wells 4
Pre-Mixed Standard 2 Standard Diluent 1×20mL
Pre-Mixed Magnetic beads (22#:IL34) 1 Analysis buffer 1×20mL
Pre-Mixed Detection Reagent A 1×120μL Assay Diluent A 1×12mL
Detection Reagent B (PE-SA) 1×120μL Assay Diluent B 1×12mL
Sheath Fluid 1×10mL Wash Buffer (30 × concentrate) 1×20mL
Instruction manual 1

Assay procedure summary

1. Preparation of standards, reagents and samples before the experiment;
2. Add 100μL standard or sample to each well,
    add 10μL magnetic beads, and incubate 90min at 37°C on shaker;
3. Remove liquid on magnetic frame, add 100μL prepared Detection Reagent A. Incubate 60min at 37°C on shaker;
4. Wash plate on magnetic frame for three times;
5. Add 100μL prepared Detection Reagent B, and incubate 30 min at 37°C on shaker;
6. Wash plate on magnetic frame for three times;
7. Add 100μL sheath solution, swirl for 2 minutes, read on the machine.

GIVEAWAYS

INCREMENT SERVICES

Magazine Citations
J Int Med Res. Increased Serum Interleukin-34 in Patients with Coronary Artery Disease PubMed: 23206468
Transplantation IL-34 inhibits acute rejection of rat liver transplantation by inducing Kupffer cell M2 polarization Pubmed:29570162
cyberleninka КЛИНИКО-ЭПИДЕМИОЛОГИЧЕСКАЯ ХАРАКТЕРИСТИКА И ОСОБЕННОСТИ БАЛЛАНСА ЦИТОКИНОВ ПРИ КЛЕЩЕВОМ ЭНЦЕФАЛИТЕ В …
Revista da Associacao Medica Brasileira A comparison of IL-17 and IL-34 concentrations in the cerebrospinal fluid of patients with acute inflammatory demyelinating neuropathy and chronic …
Rev Assoc Med Bras A comparison of IL-17 and IL-34 concentrations in the cerebrospinal fluid of patients with acute inflammatory demyelinating neuropathy and chronic?¡­ 33295414
Catalog No. Related products for research use of Homo sapiens (Human) Organism species Applications (RESEARCH USE ONLY!)
APC007Hu61 Active Interleukin 34 (IL34) Cell culture; Activity Assays.
APC007Hu01 Active Interleukin 34 (IL34) Cell culture; Activity Assays.
RPC007Hu02 Recombinant Interleukin 34 (IL34) Positive Control; Immunogen; SDS-PAGE; WB.
EPC007Hu62 Eukaryotic Interleukin 34 (IL34) Positive Control; Immunogen; SDS-PAGE; WB.
RPC007Hu01 Recombinant Interleukin 34 (IL34) Positive Control; Immunogen; SDS-PAGE; WB.
PAC007Hu01 Polyclonal Antibody to Interleukin 34 (IL34) WB; IHC; ICC; IP.
MAC007Hu21 Monoclonal Antibody to Interleukin 34 (IL34) WB; IHC; ICC; IP.
MAC007Hu23 Monoclonal Antibody to Interleukin 34 (IL34) WB; IHC; ICC; IP.
MAC007Hu22 Monoclonal Antibody to Interleukin 34 (IL34) WB; IHC; ICC; IP.
SEC007Hu ELISA Kit for Interleukin 34 (IL34) Enzyme-linked immunosorbent assay for Antigen Detection.
SCC007Hu CLIA Kit for Interleukin 34 (IL34) Chemiluminescent immunoassay for Antigen Detection.
LMC007Hu Multiplex Assay Kit for Interleukin 34 (IL34) ,etc. by FLIA (Flow Luminescence Immunoassay) FLIA Kit for Antigen Detection.