Multiplex Assay Kit for Interleukin 21 (IL21) ,etc. by FLIA (Flow Luminescence Immunoassay)

Za11

(Note: Up to 8-plex in one testing reaction)

Specificity

This assay has high sensitivity and excellent specificity for detection of Interleukin 21 (IL21) ,etc. by FLIA (Flow Luminescence Immunoassay).
No significant cross-reactivity or interference between Interleukin 21 (IL21) ,etc. by FLIA (Flow Luminescence Immunoassay) and analogues was observed.

Recovery

Matrices listed below were spiked with certain level of recombinant Interleukin 21 (IL21) ,etc. by FLIA (Flow Luminescence Immunoassay) and the recovery rates were calculated by comparing the measured value to the expected amount of Interleukin 21 (IL21) ,etc. by FLIA (Flow Luminescence Immunoassay) in samples.

Matrix Recovery range (%) Average(%)
serum(n=5) 78-91 88
EDTA plasma(n=5) 94-102 98
heparin plasma(n=5) 79-91 84

Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Interleukin 21 (IL21) ,etc. by FLIA (Flow Luminescence Immunoassay) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Interleukin 21 (IL21) ,etc. by FLIA (Flow Luminescence Immunoassay) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Interleukin 21 (IL21) ,etc. by FLIA (Flow Luminescence Immunoassay) and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Sample 1:2 1:4 1:8 1:16
serum(n=5) 86-95% 82-104% 95-102% 78-97%
EDTA plasma(n=5) 78-89% 96-103% 88-96% 79-91%
heparin plasma(n=5) 99-105% 86-98% 81-104% 78-90%

Stability

The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

Reagents and materials provided

Reagents Quantity Reagents Quantity
96-well plate 1 Plate sealer for 96 wells 4
Pre-Mixed Standard 2 Standard Diluent 1×20mL
Pre-Mixed Magnetic beads (22#:IL21) 1 Analysis buffer 1×20mL
Pre-Mixed Detection Reagent A 1×120μL Assay Diluent A 1×12mL
Detection Reagent B (PE-SA) 1×120μL Assay Diluent B 1×12mL
Sheath Fluid 1×10mL Wash Buffer (30 × concentrate) 1×20mL
Instruction manual 1

Assay procedure summary

1. Preparation of standards, reagents and samples before the experiment;
2. Add 100μL standard or sample to each well,
    add 10μL magnetic beads, and incubate 90min at 37°C on shaker;
3. Remove liquid on magnetic frame, add 100μL prepared Detection Reagent A. Incubate 60min at 37°C on shaker;
4. Wash plate on magnetic frame for three times;
5. Add 100μL prepared Detection Reagent B, and incubate 30 min at 37°C on shaker;
6. Wash plate on magnetic frame for three times;
7. Add 100μL sheath solution, swirl for 2 minutes, read on the machine.

GIVEAWAYS

INCREMENT SERVICES

Magazine Citations
Neurological Sciences Serum IL-21 levels decrease with glucocorticoid treatment in myasthenia gravis Pubmed: 23708911
Physiological Research Pre-Press Article Decrease in serum interleukin 21 levels is associated with disease activity improvement in patients with recent-onset rheumatoid arthritis Biomed: Source
BioMed Research International Indoleamine 2, 3-Dioxygenase (IDO) Regulates Th17/Treg Immunity in Experimental IgA Nephropathy
Am J Transplant Circulatory follicular helper T lymphocytes associate with lower incidence of CMV infection in kidney transplant recipients 34153157
J Periodontol Cytokine profile in serum and gingival crevicular fluid of children with inflammatory bowel disease: A case‐control study 34730850
Catalog No. Related products for research use of Mus musculus (Mouse) Organism species Applications (RESEARCH USE ONLY!)
RPB688Mu02 Recombinant Interleukin 21 (IL21) Positive Control; Immunogen; SDS-PAGE; WB.
RPB688Mu01 Recombinant Interleukin 21 (IL21) Positive Control; Immunogen; SDS-PAGE; WB.
PAB688Mu02 Polyclonal Antibody to Interleukin 21 (IL21) WB; IHC; ICC; IP.
PAB688Mu01 Polyclonal Antibody to Interleukin 21 (IL21) WB; IHC; ICC; IP.
SEB688Mu ELISA Kit for Interleukin 21 (IL21) Enzyme-linked immunosorbent assay for Antigen Detection.
MEB688Mu Mini Samples ELISA Kit for Interleukin 21 (IL21) Enzyme-linked immunosorbent assay for Antigen Detection.
SCB688Mu CLIA Kit for Interleukin 21 (IL21) Chemiluminescent immunoassay for Antigen Detection.
LMB688Mu Multiplex Assay Kit for Interleukin 21 (IL21) ,etc. by FLIA (Flow Luminescence Immunoassay) FLIA Kit for Antigen Detection.