Multiplex Assay Kit for Oncostatin M (OSM) ,etc. by FLIA (Flow Luminescence Immunoassay)

OS-M

(Note: Up to 8-plex in one testing reaction)

Specificity

This assay has high sensitivity and excellent specificity for detection of Oncostatin M (OSM) ,etc. by FLIA (Flow Luminescence Immunoassay).
No significant cross-reactivity or interference between Oncostatin M (OSM) ,etc. by FLIA (Flow Luminescence Immunoassay) and analogues was observed.

Recovery

Matrices listed below were spiked with certain level of recombinant Oncostatin M (OSM) ,etc. by FLIA (Flow Luminescence Immunoassay) and the recovery rates were calculated by comparing the measured value to the expected amount of Oncostatin M (OSM) ,etc. by FLIA (Flow Luminescence Immunoassay) in samples.

Matrix Recovery range (%) Average(%)
serum(n=5) 88-104 97
EDTA plasma(n=5) 85-92 89
heparin plasma(n=5) 78-94 81

Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Oncostatin M (OSM) ,etc. by FLIA (Flow Luminescence Immunoassay) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Oncostatin M (OSM) ,etc. by FLIA (Flow Luminescence Immunoassay) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Oncostatin M (OSM) ,etc. by FLIA (Flow Luminescence Immunoassay) and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Sample 1:2 1:4 1:8 1:16
serum(n=5) 80-99% 86-98% 93-101% 92-102%
EDTA plasma(n=5) 85-92% 95-102% 78-94% 94-101%
heparin plasma(n=5) 99-105% 83-95% 80-92% 90-101%

Stability

The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

Reagents and materials provided

Reagents Quantity Reagents Quantity
96-well plate 1 Plate sealer for 96 wells 4
Pre-Mixed Standard 2 Standard Diluent 1×20mL
Pre-Mixed Magnetic beads (22#:OSM) 1 Analysis buffer 1×20mL
Pre-Mixed Detection Reagent A 1×120μL Assay Diluent A 1×12mL
Detection Reagent B (PE-SA) 1×120μL Assay Diluent B 1×12mL
Sheath Fluid 1×10mL Wash Buffer (30 × concentrate) 1×20mL
Instruction manual 1

Assay procedure summary

1. Preparation of standards, reagents and samples before the experiment;
2. Add 100μL standard or sample to each well,
    add 10μL magnetic beads, and incubate 90min at 37°C on shaker;
3. Remove liquid on magnetic frame, add 100μL prepared Detection Reagent A. Incubate 60min at 37°C on shaker;
4. Wash plate on magnetic frame for three times;
5. Add 100μL prepared Detection Reagent B, and incubate 30 min at 37°C on shaker;
6. Wash plate on magnetic frame for three times;
7. Add 100μL sheath solution, swirl for 2 minutes, read on the machine.

GIVEAWAYS

INCREMENT SERVICES

Magazine Citations
Journal of Obstetrics and Gynaecology Research Oncostatin M as a target biological molecule of preeclampsia PubMed: 20149034
J Neuroinflammation Oncostatin M promotes excitotoxicity by inhibiting glutamate uptake in astrocytes: implications in HIV-associated neurotoxicity via Smad2/4 in Human Renal Proximal Tubular Epithelial Cells Pubmed:27287400
22 Modulating the phenotype of host macrophages to enhance osteogenesis in MSC-laden hydrogels: Design of a glucomannan coating material. pubmed:28582717
Advanced Functional Materials Fungal Component Coating Enhances Titanium Implant‐Bone Integration Doi: 10.1002/adfm.201804483
Catalog No. Related products for research use of Homo sapiens (Human) Organism species Applications (RESEARCH USE ONLY!)
EPA110Hu61 Eukaryotic Oncostatin M (OSM) Positive Control; Immunogen; SDS-PAGE; WB.
RPA110Hu01 Recombinant Oncostatin M (OSM) Positive Control; Immunogen; SDS-PAGE; WB.
APA110Hu01 Active Oncostatin M (OSM) Cell culture; Activity Assays.
PAA110Hu06 Polyclonal Antibody to Oncostatin M (OSM) WB; IHC; ICC; IP.
PAA110Hu01 Polyclonal Antibody to Oncostatin M (OSM) WB; IHC; ICC; IP.
MAA110Hu22 Monoclonal Antibody to Oncostatin M (OSM) WB; IHC; ICC; IP.
SEA110Hu ELISA Kit for Oncostatin M (OSM) Enzyme-linked immunosorbent assay for Antigen Detection.
SCA110Hu CLIA Kit for Oncostatin M (OSM) Chemiluminescent immunoassay for Antigen Detection.
LMA110Hu Multiplex Assay Kit for Oncostatin M (OSM) ,etc. by FLIA (Flow Luminescence Immunoassay) FLIA Kit for Antigen Detection.
KSA110Hu01 ELISA Kit DIY Materials for Oncostatin M (OSM) Main materials for "Do It (ELISA Kit) Yourself".