CLIA Kit for Immunoglobulin A2 (IgA2)

IGHA2; Immunoglobulin Heavy Constant Alpha 2; A2m Marker

Specificity

This assay has high sensitivity and excellent specificity for detection of Immunoglobulin A2 (IgA2).
No significant cross-reactivity or interference between Immunoglobulin A2 (IgA2) and analogues was observed.

Recovery

Matrices listed below were spiked with certain level of recombinant Immunoglobulin A2 (IgA2) and the recovery rates were calculated by comparing the measured value to the expected amount of Immunoglobulin A2 (IgA2) in samples.

Matrix Recovery range (%) Average(%)
serum(n=5) 81-101 95
EDTA plasma(n=5) 98-105 101
heparin plasma(n=5) 92-101 97

Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Immunoglobulin A2 (IgA2) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Immunoglobulin A2 (IgA2) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Immunoglobulin A2 (IgA2) and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Sample 1:2 1:4 1:8 1:16
serum(n=5) 82-101% 79-99% 86-101% 87-97%
EDTA plasma(n=5) 92-101% 84-96% 91-99% 80-102%
heparin plasma(n=5) 79-97% 81-97% 90-98% 85-99%

Stability

The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

Reagents and materials provided

Reagents Quantity Reagents Quantity
Pre-coated, ready to use 96-well strip plate 1 Plate sealer for 96 wells 4
Standard 2 Standard Diluent 1×20mL
Detection Reagent A 1×120µL Assay Diluent A 1×12mL
Detection Reagent B 1×120µL Assay Diluent B 1×12mL
Substrate A 1×10mL Substrate B 1×2mL
Wash Buffer (30 × concentrate) 1×20mL Instruction manual 1

Assay procedure summary

1. Prepare all reagents, samples and standards;
2. Add 100µL standard or sample to each well. Incubate 1 hours at 37°C;
3. Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C;
4. Aspirate and wash 3 times;
5. Add 100µL prepared Detection Reagent B. Incubate 30 minutes at 37°C;
6. Aspirate and wash 5 times;
7. Add 100µL Substrate Solution. Incubate 10 minutes at 37°C;
8. Read RLU value immediately.

GIVEAWAYS

INCREMENT SERVICES

Catalog No. Related products for research use of Homo sapiens (Human) Organism species Applications (RESEARCH USE ONLY!)
NPE777Hu01 Native Immunoglobulin A2 (IgA2) Positive Control; Immunogen; SDS-PAGE; WB.
RPE777Hu01 Recombinant Immunoglobulin A2 (IgA2) Positive Control; Immunogen; SDS-PAGE; WB.
PAE777Hu01 Polyclonal Antibody to Immunoglobulin A2 (IgA2) WB; IHC; ICC; IP.
MAE777Hu21 Monoclonal Antibody to Immunoglobulin A2 (IgA2) WB; IHC; ICC; IP.
SEE777Hu ELISA Kit for Immunoglobulin A2 (IgA2) Enzyme-linked immunosorbent assay for Antigen Detection.
SCE777Hu CLIA Kit for Immunoglobulin A2 (IgA2) Chemiluminescent immunoassay for Antigen Detection.
LME777Hu Multiplex Assay Kit for Immunoglobulin A2 (IgA2) ,etc. by FLIA (Flow Luminescence Immunoassay) FLIA Kit for Antigen Detection.