CLIA Kit for Fibroblast Growth Factor 18 (FGF18)

ZFGF5

Specificity

This assay has high sensitivity and excellent specificity for detection of Fibroblast Growth Factor 18 (FGF18).
No significant cross-reactivity or interference between Fibroblast Growth Factor 18 (FGF18) and analogues was observed.

Recovery

Matrices listed below were spiked with certain level of recombinant Fibroblast Growth Factor 18 (FGF18) and the recovery rates were calculated by comparing the measured value to the expected amount of Fibroblast Growth Factor 18 (FGF18) in samples.

Matrix Recovery range (%) Average(%)
serum(n=5) 97-105 101
EDTA plasma(n=5) 82-97 87
heparin plasma(n=5) 93-101 97

Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Fibroblast Growth Factor 18 (FGF18) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Fibroblast Growth Factor 18 (FGF18) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Fibroblast Growth Factor 18 (FGF18) and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Sample 1:2 1:4 1:8 1:16
serum(n=5) 96-103% 92-103% 97-105% 87-103%
EDTA plasma(n=5) 84-91% 90-97% 88-96% 80-103%
heparin plasma(n=5) 83-99% 94-105% 79-98% 95-105%

Stability

The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

Reagents and materials provided

Reagents Quantity Reagents Quantity
Pre-coated, ready to use 96-well strip plate 1 Plate sealer for 96 wells 4
Standard 2 Standard Diluent 1×20mL
Detection Reagent A 1×120µL Assay Diluent A 1×12mL
Detection Reagent B 1×120µL Assay Diluent B 1×12mL
Substrate A 1×10mL Substrate B 1×2mL
Wash Buffer (30 × concentrate) 1×20mL Instruction manual 1

Assay procedure summary

1. Prepare all reagents, samples and standards;
2. Add 100µL standard or sample to each well. Incubate 1 hours at 37°C;
3. Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C;
4. Aspirate and wash 3 times;
5. Add 100µL prepared Detection Reagent B. Incubate 30 minutes at 37°C;
6. Aspirate and wash 5 times;
7. Add 100µL Substrate Solution. Incubate 10 minutes at 37°C;
8. Read RLU value immediately.

GIVEAWAYS

INCREMENT SERVICES

Magazine Citations
Pesqui. Bras. Odontopediatria Clín. Integr Human Umbilical Cord Mesenchymal Stem Cells Over Platelet Rich Fibrin Scaffold for Mandibular Cartilage Defects Regenerative Medicine
J Exp Clin Cancer Res HDAC7 promotes NSCLC proliferation and metastasis via stabilization by deubiquitinase USP10 and activation of β-catenin-FGF18 pathway Pubmed:35277183
Catalog No. Related products for research use of Rattus norvegicus (Rat) Organism species Applications (RESEARCH USE ONLY!)
RPC907Ra01 Recombinant Fibroblast Growth Factor 18 (FGF18) Positive Control; Immunogen; SDS-PAGE; WB.
PAC907Ra01 Polyclonal Antibody to Fibroblast Growth Factor 18 (FGF18) WB; IHC; ICC; IP.
MAC907Ra21 Monoclonal Antibody to Fibroblast Growth Factor 18 (FGF18) WB; IHC; ICC; IP.
SEC907Ra ELISA Kit for Fibroblast Growth Factor 18 (FGF18) Enzyme-linked immunosorbent assay for Antigen Detection.
SCC907Ra CLIA Kit for Fibroblast Growth Factor 18 (FGF18) Chemiluminescent immunoassay for Antigen Detection.
LMC907Ra Multiplex Assay Kit for Fibroblast Growth Factor 18 (FGF18) ,etc. by FLIA (Flow Luminescence Immunoassay) FLIA Kit for Antigen Detection.